Radiation exposure to the hands from mini C-arm fluoroscopy.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Singer Gordon
Аннотация
PURPOSE: To quantify the level of radiation exposure to the hands of hand surgeons using intraoperative mini C-arm fluoroscopy and to compare the actual level of exposure with predicted levels and acceptable limits.
METHODS: Five hand surgeons were given ring dosimeters to measure radiation exposure to their hands during surgery of the finger, hand, and wrist. A total of 81 rings were analyzed. After the clinical study a phantom was used to measure scatter at close range from the mini C-arm.
RESULTS: Surgeons' hands were exposed to an average +/- SD of 20 +/- 12.3 mrem/case. For comparison a chest x-ray results in approximately 20 mrem exposure to the patient. Radiation exposure for the group of hand surgeons ranged from 5 to 80 mrem. Surgeons used an average of 51 +/- 36.9 seconds of fluoroscopy time per case. Exposure time for the group ranged from 6 to 170 seconds. The radiation scatter rate decreases precipitously outside the beam or beyond the radius of the intensifier. An average exposure to the hands of 20 mrem/case suggests that surgeons' hands must be entering the beam and getting direct exposure.
CONCLUSIONS: Hand surgeons work close to the beam and as a result their hands potentially are exposed to a nontrivial amount of radiation. We recommend that surgeons who use the mini C-arm use precautions to minimize radiation exposure, particularly to their hands.
METHODS: Five hand surgeons were given ring dosimeters to measure radiation exposure to their hands during surgery of the finger, hand, and wrist. A total of 81 rings were analyzed. After the clinical study a phantom was used to measure scatter at close range from the mini C-arm.
RESULTS: Surgeons' hands were exposed to an average +/- SD of 20 +/- 12.3 mrem/case. For comparison a chest x-ray results in approximately 20 mrem exposure to the patient. Radiation exposure for the group of hand surgeons ranged from 5 to 80 mrem. Surgeons used an average of 51 +/- 36.9 seconds of fluoroscopy time per case. Exposure time for the group ranged from 6 to 170 seconds. The radiation scatter rate decreases precipitously outside the beam or beyond the radius of the intensifier. An average exposure to the hands of 20 mrem/case suggests that surgeons' hands must be entering the beam and getting direct exposure.
CONCLUSIONS: Hand surgeons work close to the beam and as a result their hands potentially are exposed to a nontrivial amount of radiation. We recommend that surgeons who use the mini C-arm use precautions to minimize radiation exposure, particularly to their hands.