MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00256-016-2350-6

Медицинская библиотека · Skeletal radiology · 2016

Comparison of 2D radiography and a semi-automatic CT-based 3D method for measuring change in dorsal angulation over time in distal radius fractures.

Christersson Albert, Nysjö Johan, Berglund Lars, Malmberg Filip, Sintorn Ida-Maria, Nyström Ingela, Larsson Sune

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

The aim of the present study was to compare the reliability and agreement between a computer tomography-based method (CT) and digitalised 2D radiographs (XR) when measuring change in dorsal angulation over time in distal radius fractures.

MATERIALS AND

METHODS

Radiographs from 33 distal radius fractures treated with external fixation were retrospectively analysed. All fractures had been examined using both XR and CT at six times over 6 months postoperatively. The changes in dorsal angulation between the first reference images and the following examinations in every patient were calculated from 133 follow-up measurements by two assessors and repeated at two different time points. The measurements were analysed using Bland-Altman plots, comparing intra- and inter-observer agreement within and between XR and CT.

RESULTS

The mean differences in intra- and inter-observer measurements for XR, CT, and between XR and CT were close to zero, implying equal validity. The average intra- and inter-observer limits of agreement for XR, CT, and between XR and CT were ± 4.4°, ± 1.9° and ± 6.8° respectively.

CONCLUSIONS

For scientific purpose, the reliability of XR seems unacceptably low when measuring changes in dorsal angulation in distal radius fractures, whereas the reliability for the semi-automatic CT-based method was higher and is therefore preferable when a more precise method is requested.

Для цитирования: Christersson Albert, Nysjö Johan, Berglund Lars, Malmberg Filip, Sintorn Ida-Maria, Nyström Ingela, Larsson Sune. Comparison of 2D radiography and a semi-automatic CT-based 3D method for measuring change in dorsal angulation over time in distal radius fractures.. Skeletal radiology. 2016. DOI: 10.1007/s00256-016-2350-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.