Аннотация

BACKGROUND

Edwards Intuity is designed for rapid deployment based on the structure of Magna Ease. This study was conducted to compare early hemodynamic performance between the two valves.

METHODS

Patients who underwent aortic valve replacement (AVR) using Edwards Intuity or Carpentier-Edwards PERIMOUNT Magna Ease in our institution from June 2016 to July 2021 were enrolled. Intuity valve was used in 215 patients, and Magna Ease valve was used in 198 patients, respectively. Early postoperative echocardiographic data were available in 99.0% (409/413) of the patients. The transvalvular mean pressure gradient, effective orifice area, and effective orifice area index were compared between the valves stratified by prosthesis size.

RESULTS

There were no differences in the proportion of female patients or body surface area between the groups. Mean pressure gradient on early postoperative echocardiography was significantly lower in Intuity valve than Magna Ease valve for 19, 21, 23, and 25 mm valves (15.5±5.0 vs. 20.8±9.1 mmHg, P=0.004; 12.7±4.2 vs. 15.6±5.3 mmHg, P=0.001; 11.5±3.3 vs. 13.4±5.8 mmHg, P=0.034; and 9.9±3.1 vs. 12.3±4.0 mmHg, P=0.029; respectively). Effective orifice area was larger in Intuity valve than Magna Ease valve for 19 mm valve (1.45±0.38 vs. 1.19±0.28 cm2, P=0.002), and effective orifice area index was also larger in Intuity valve than Magna Ease valve for 19 mm valve (0.96±0.26 vs. 0.80±0.20 cm2/m2, P=0.005). Early clinical outcomes, including operative mortality and postoperative complications, demonstrated no significant differences between the groups.

CONCLUSIONS

Edwards Intuity demonstrated superior early hemodynamic performance compared with Magna Ease in a size-by-size comparison, and this superiority was more definite for small prostheses.

Для цитирования: Sohn Suk Ho, Kim Kyung Hwan, Kang Yoonjin, Kim Ji Seong, Choi Jae Woong, Hwang Ho Young. Rapid deployment versus its conventional counterpart in aortic valve replacement: comparison of early hemodynamic outcomes.. Journal of thoracic disease. 2023. DOI: 10.21037/jtd-23-318.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.