MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.28360

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2019

Long-term transesophageal echocardiography after patent foramen ovale closure by BioSTAR and Amplatzer patent foramen ovale occluders.

Suligoj Natasa Cernic, Rojko Maja, Suligoj Brina, Zorc Metka, Kar Saibal, Noc Marko

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

To compare transesophageal echocardiography (TOE) findings after patent foramen ovale (PFO) closure by BioSTAR (NMT Medical Inc, Boston, MA) and Amplatzer PFO occluders (Abbott Vascular, Plymouth, MN).

BACKGROUND

PFO closure with a biodegradable device represents an attractive alternative to permanent devices. Long-term effectiveness and morphology after biodegradation remain unknown.

METHODS

Between February 2008 and June 2014, 49 patients received BioSTAR and 48 Amplatzer PFO occluder. TOE was performed after closure, at 6 months and beyond 2 years.

RESULTS

PFO features were comparable between the groups. Immediate effective closure (<5 bubbles on Valsalva) was obtained in 96% by BioSTAR and 88% by Amplatzer PFO occluder (p = .16). Except for transient fever after BioSTAR (10.8 vs. 0%; p = .08), there was no adverse events. TOE at 6 months revealed comparable effective closure (93 vs. 89%; p = .74), all devices in correct position and no thrombus/pericardial effusion. In the BioSTAR group, a peri-device left-to-right color Doppler shunt was documented in one patient (2.2%), protrusion of the nitinol framework strut(s) into the atrial cavity in two patients (4.3%), and both events in one patient (2.2%). TOE beyond 2 years showed comparable effective closure (92 vs. 96%; p = 1.00) and again BioSTAR-associated peri-device left-to-right shunt and metal framework strut(s) protrusion. There was no stroke or peripheral embolization in either group while TIA was numerically greater in BioSTAR patients (6.8 vs. 2.5%; p = .61).

CONCLUSION

BioSTAR provided similar PFO closure rate as Amplatzer PFO occluder. As yet unreported BioSTAR-associated peri-device left-to-right shunt and metal framework strut(s) protrusion may have practical implications for further development of biodegradable devices.

Для цитирования: Suligoj Natasa Cernic, Rojko Maja, Suligoj Brina, Zorc Metka, Kar Saibal, Noc Marko. Long-term transesophageal echocardiography after patent foramen ovale closure by BioSTAR and Amplatzer patent foramen ovale occluders.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2019. DOI: 10.1002/ccd.28360.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.