Isotopic Scintigraphy Coupled With Computed Tomography for the Investigation of Intrathecal Baclofen Device Malfunction.

Frémondière Flavie, Lacoeuille Franck, Sher Avigaelle, Couturier Olivier, Menei Philippe, Richard Isabelle, Dinomais Mickaël

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To assess the potential use of indium-111 diethylenetriamine pentaacetic acid ((111)In-DTPA) scintigraphy coupled with computed tomography (CT) for the investigation of intrathecal baclofen (ITB) device malfunction.

DESIGN

Retrospective study of a case series of patients.

SETTING

Neurosurgical and physical and rehabilitation medicine departments.

PARTICIPANTS

Patients (N=7) with reduced ITB effectiveness in whom prior conventional radiographs were inconclusive.

INTERVENTION: Nine (111)In-DTPA scintigraphic studies and 8 CT scans. Planar acquisitions were followed by tomoscintigraphy combined with CT.

MAIN OUTCOME MEASURE: Progression of the radiotracer in the pump, catheters, and in the subarachnoid space.

RESULTS

In 7 cases, scintigraphy coupled with CT showed leakage behind the pump, lack of activity outside the pump reservoir, abrupt interruption of activity in the catheter, or abnormal distribution of the radiotracer, thus demonstrating that the drug did not reach its target. Surgical revision confirmed these findings in 5 cases. In 1 case, combined imagery ruled out device dysfunction. In the remaining case, only planar acquisitions were performed, showing correct diffusion of the radiotracer.

CONCLUSIONS

The combination of scintigraphy and CT provides simultaneous functional and anatomic imagery of the device. The slow infusion of the radioisotope mimics the diffusion of baclofen, and this could be a useful method to explore intrathecal device malfunction. Further studies are required to compare scintigraphy coupled with CT, to radiopaque injection followed by fluoroscopy or CT.

Для цитирования: Frémondière Flavie, Lacoeuille Franck, Sher Avigaelle, Couturier Olivier, Menei Philippe, Richard Isabelle, Dinomais Mickaël. Isotopic Scintigraphy Coupled With Computed Tomography for the Investigation of Intrathecal Baclofen Device Malfunction.. Archives of physical medicine and rehabilitation. 2015. DOI: 10.1016/j.apmr.2015.10.101.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.