Temporary balloon occlusion of atrial septal defects in suspected or documented left ventricular diastolic dysfunction: Hemodynamic and clinical findings.
Аннотация
OBJECTIVE
To review our experience with balloon testing prior to atrial septal defect (ASD) closure in adults with left ventricular (LV) diastolic dysfunction.
BACKGROUND
ASD closure in patients with LV diastolic dysfunction may precipitate LV failure. Temporary ASD occlusion has been used in this scenario but data are limited.
METHODS
Retrospective review of 27 patients age ≥ 50 years undergoing temporary ASD balloon occlusion between 2000 and 2018 for suspected LV diastolic dysfunction or elevated LV end-diastolic pressure (LVEDP).
RESULTS
Median age was 72 years (IQR 66.7; 75.2). Atrial fibrillation was seen in 58% of patients, hypertension in 58%, and coronary artery disease in 26%; 52% were females. Median ASD size was 13 mm (10; 18) and Qp/Qs 1.8 (1.6; 2.2). Median LVEDP was 14 mmHg (12; 22); pulmonary artery wedge pressure (PAWP) 12 mmHg (9; 16.5) and left atrial pressure (LAP) 13.5 mmHg (8; 16.3). After a median of 5 min (3; 10) of balloon occlusion, patients with baseline LVEDP ≥15 mmHg had more significant increases in LVEDP (9 [6; 12] vs. 2 mmHg [0.5; 5]; p = 0.03) and LAP/PAWP (10.5 [8.3; 16.3] vs. 1.5 mmHg [-1.5; 3]; p = 0.0003) than those with baseline LVEDP <15 mmHg. None of those with a baseline LVEDP <15 mmHg had a LAP/PAWP >15 mmHg during balloon testing compared to 92% of patients with a baseline LVEDP ≥15 mmHg.
CONCLUSION
LVEDP might be used to predict LAP post-ASD closure. Comorbidities typically associated with LV diastolic dysfunction are common in these patients and should be considered in their management.