DOI 10.1007/s00134-011-2355-5

Медицинская библиотека · Intensive care medicine · 2011

Delivery of helium–oxygen mixture during spontaneous breathing: evaluation of three high-concentration face masks.

Roche-Campo Ferran, Vignaux Laurence, Galia Fabrice, Lyazidi Aissam, Vargas Frédéric, Texereau Joëlle, Apiou-Sbirlea Gabriela, Jolliet Philippe, Brochard Laurent

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To evaluate the efficacy of delivering a mixture of helium and oxygen gas (He–O2) in spontaneous ventilation. Three high oxygen flow reservoir masks were tested: the Heliox21, specifically designed for helium; the Hi-Ox80 mask, with an inspiratory and an expiratory valve; and a standard high-concentration face mask.

METHODS

This prospective randomized crossover study was performed in six healthy volunteers in a laboratory setting. Volunteers breathed a mixture of 78% He/22% O2 through each of the masks under two different breathing conditions (rest and hyperventilation: minute ventilation of 14.9 ± 6.1 and 26.7 ± 8.7 L min(−1), respectively) and four different He–O2 flow rates (7, 10, 12, and 15 L min(−1)).

RESULTS

A nasopharyngeal catheter was used to estimate He pharyngeal concentration (Fp [He]) in the airways in order to determine the percentage of contamination with room air (% air cont) at end-expiration. Under all testing conditions, the Hi-Ox80 mask presented a significantly lower % air cont. During resting breathing pattern, a Fp [He] higher than 50% was achieved in 54% of the tests performed with the Hi-Ox80 mask compared to 29% for the Heliox21 mask and only 17% for the standard mask. At hyperventilation, a Fp [He] higher than 50% was achieved in 17% of the tests performed with the Hi-Ox mask compared to 4% for the other two masks.

CONCLUSION

He–O2 administration via the usual high-concentration reservoir masks results in significant dilution by room air. The Hi-Ox80 mask minimized room air contamination and much more frequently achieved a pharyngeal He concentration higher than 50%.

Для цитирования: Roche-Campo Ferran, Vignaux Laurence, Galia Fabrice, Lyazidi Aissam, Vargas Frédéric, Texereau Joëlle, Apiou-Sbirlea Gabriela, Jolliet Philippe, Brochard Laurent. Delivery of helium–oxygen mixture during spontaneous breathing: evaluation of three high-concentration face masks.. Intensive care medicine. 2011. DOI: 10.1007/s00134-011-2355-5.