Late Balloon Valvuloplasty for Transcatheter Heart Valve Dysfunction.

Akodad Mariama, Blanke Philipp, Chuang Ming-Yu A, Duchscherer Jade, Sellers Stephanie L, Chatfield Andrew G, Gulsin Gaurav G, Lauck Sandra, Leipsic Jonathon A, Meier David, Moss Rob R, Cheung Anson, Sathananthan Janarthanan, Wood David A, Ye Jian, Webb John G

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Transcatheter heart valve (THV) dysfunction with an elevated gradient or paravalvular leak (PVL) may be documented late after THV implantation. Medical management, paravalvular plugs, redo THV replacement, or surgical valve replacement may be considered. However, late balloon dilatation is rarely utilized because of concerns about safety or lack of efficacy.

OBJECTIVES

We aimed to evaluate the safety and efficacy of late dilatation in the management of THV dysfunction.

METHODS

All patients who underwent late dilatation for symptomatic THV dysfunction at 2 institutions between 2016 and 2021 were identified. Baseline, procedural characteristics, and clinical and echocardiographic outcomes were documented. THV frame expansion was assessed by multislice computed tomography before and after late dilatation.

RESULTS

Late dilatation was performed in 30 patients a median of 4.6 months (IQR: 2.3-11.0 months) after THV implantation in the aortic (n = 25; 83.3%), mitral (n = 2; 6.7%), tricuspid (n = 2; 6.7%) and pulmonary (n = 1; 3.3%) position. THV underexpansion was documented at baseline, and frame expansion substantially improved after late dilatation. The mean transvalvular gradient fell in all patients. For aortic THVs specifically, mean transaortic gradient fell from 25.4 ± 13.9 mm Hg to 10.8 ± 4.1 mm Hg; P < 0.001. PVL was reduced to ≤mild in all 11 patients with a previous >mild PVL. Embolic events, stroke, annular injury, and bioprosthetic leaflet injury were not observed. Symptomatic benefit was durable at 19.6 months (IQR: 14.8-36.1 months) follow-up.

CONCLUSIONS

Balloon dilatation late after THV implantation appears feasible and safe in appropriately selected patients and may result in THV frame expansion resulting in improvements in hemodynamic performance and PVL.

Для цитирования: Akodad Mariama, Blanke Philipp, Chuang Ming-Yu A, Duchscherer Jade, Sellers Stephanie L, Chatfield Andrew G, Gulsin Gaurav G, Lauck Sandra, Leipsic Jonathon A, Meier David, Moss Rob R, Cheung Anson, Sathananthan Janarthanan, Wood David A, Ye Jian, Webb John G. Late Balloon Valvuloplasty for Transcatheter Heart Valve Dysfunction.. Journal of the American College of Cardiology. 2022. DOI: 10.1016/j.jacc.2022.01.041.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.