Absence of electrocardiographic change after prolonged application of a conducted electrical weapon in physically exhausted adults.

Ho Jeffrey D, Dawes Donald M, Heegaard William G, Calkins Hugh G, Moscati Ronald M, Miner James R

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Conducted electrical weapons (CEWs) are used by law enforcement for control of subjects by causing neuromuscular incapacitation. There has been scrutiny of CEWs and their potential role in the occasional sudden death of subjects in custody. There is a hypothesized causal relationship due to induced cardiac dysrhythmia. Previous work has not shown dysrhythmia induction in resting humans. However, these devices are not often used on resting individuals in the field.

OBJECTIVE

We sought to determine if exposure to a CEW in a physically exhausted human sample population caused detectable change in the 12-lead electrocardiogram (ECG).

METHODS

Human volunteers were enrolled. All subjects had a baseline ECG obtained and then underwent an exercise regimen until exhaustion. The volunteers then received a continuous 15-s application from a TASER® X26 CEW (TASER International, Scottsdale, AZ). CEW electrodes were placed on random positions of their anterior thoraces. Electrode positions involved at least a 12-inch spread and always encompassed the normal anatomic position of the heart. An ECG was obtained immediately after CEW exposure. ECGs were interpreted by a blinded cardiologist.

RESULTS

At baseline, 24/25 ECGs were normal. One baseline ECG was abnormal due to several monomorphic premature ventricular complexes. After CEW exposure, all 25 ECGs were interpreted as normal.

CONCLUSIONS

Prolonged CEW application in an exhausted human sample did not cause a detectable change in their 12-lead ECGs. Theories of CEW-induced dysrhythmia in non-rested humans are not supported by our findings.

Для цитирования: Ho Jeffrey D, Dawes Donald M, Heegaard William G, Calkins Hugh G, Moscati Ronald M, Miner James R. Absence of electrocardiographic change after prolonged application of a conducted electrical weapon in physically exhausted adults.. The Journal of emergency medicine. 2009. DOI: 10.1016/j.jemermed.2009.03.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.