DOI 10.1186/s12871-023-02263-1

Медицинская библиотека · BMC anesthesiology · 2023

Comparison of the effect of pressure control and volume control ventilation on endotracheal tube cuff pressure in patients undergoing general anesthesia and mechanical ventilation: a parallel randomized clinical trial.

Nasrolahzadeh Shahram, Nourian Javad, Khosravi Ahmad, Ghasempour Saeed, Abbasi Ali, Ebrahimi Hossein

Рандомизированное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Endotracheal intubation and mechanical ventilation are prevalent interventions in the operating room and intensive care unit. Recently, the complications of endotracheal tube cuff pressure have been a topic of interest. Therefore, this study compared the effect of pressure control and volume control ventilation modes on the endotracheal cuff pressure rate in patients undergoing general anesthesia and mechanical ventilation.

METHODS

In this triple-blinded randomized clinical trial, 50 patients undergoing open limb surgery and inguinal hernia were allocated to two groups of 25 based on inclusion criteria. After intubation, one group underwent ventilation on the pressure control ventilation mode, and the other underwent ventilation on the volume control ventilation mode. In both groups, using a manometer, the cuff's pressure was first adjusted in the range of 25-30 cm of water. Then, the cuff pressure was measured at 10, 20, and 30 min intervals. The data were statistically analyzed using independent t-test, and two-way repeated measures ANOVA.

RESULTS

The present study's findings showed that cuff pressure has significantly decreased over time in both study groups (F = 117.7, P < 0.001). However, a repeated measures ANOVA with a Greenhouse-Geisser correction showed no interaction between time and groups (F = 0.019, P = 0.98). The two groups had no significant difference in cuff pressure (F = 0.56, P = 0.458).

CONCLUSION

Since the cuff pressure has been significantly reduced in both groups over time, continuous monitoring of endotracheal tube cuff pressure in patients undergoing mechanical ventilation is essential. Therefore, it is suggested to keep the cuff pressure within the recommended range to prevent complications resulting from cuff pressure reduction, such as aspiration and ventilation decrease.

TRIAL REGISTRATION: The study was registered in the Iranian Registry of Clinical Trial on 23/02/2019 (trial registration number: IRCT20181018041376N1).

Для цитирования: Nasrolahzadeh Shahram, Nourian Javad, Khosravi Ahmad, Ghasempour Saeed, Abbasi Ali, Ebrahimi Hossein. Comparison of the effect of pressure control and volume control ventilation on endotracheal tube cuff pressure in patients undergoing general anesthesia and mechanical ventilation: a parallel randomized clinical trial.. BMC anesthesiology. 2023. DOI: 10.1186/s12871-023-02263-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.