DOI 10.3760/cma.j.cn121094-20200603-00317
[Determination of Methoxyacetic acid in urine by pre-column derivatization-liquid-liquid microextraction coupled with gas chromatography].
Аннотация
Objective
To establish a method for determining methoxyacetic acid in urine by pre-column derivatization-liquid-liquid microextraction coupled with gas chromatography (GC) .
Methods
Phosphate buffer solution, tert-butoxyacetic acid (internal standard) and pentafluorobenzyl bromide (derivative) were added to the urine sample. After derived in a water bath at 90 ℃ for 40 min, the mixture was cooled and filtered, then the dichloromethane was used as an extractant. After being shaken and centrifuged, the lower organic phase was sucked and injected into a gas chromatograph, separated by a DB-5 capillary column, and detected by an ECD detector.
Results
The linear range of the method was 0.6~60.0 mg/L with the correlation coefficients (r) above 0.999. The average recovery was76.6%~110.7%, the inter-day precision was 8.00%~8.82%, and the detection limit was 0.13 mg/L.
Conclusion
The method was founded to be high sensitivity, low organic reagent usage and green. So it is suitable for the detection of methoxyacetic acid in urine of occupational exposure to ethylene glycol monomethyl ether.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.