MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/jclinpath-2021-207863

Медицинская библиотека · Journal of clinical pathology · 2021

Performance of the CellaVision DC-1 digital cell imaging analyser for differential counting and morphological classification of blood cells.

van der Vorm Lisa N, Hendriks Henriët A, Smits Simone M

СтатьяПеревод в очереди

Аннотация

AIMS: Recently, a new automated digital cell imaging analyser (Sysmex CellaVision DC-1), intended for use in low-volume and small satellite laboratories, has become available. The purpose of this study was to compare the performance of the DC-1 with the Sysmex DI-60 system and the gold standard, manual microscopy.

METHODS

White blood cell (WBC) differential counts in 100 normal and 100 abnormal peripheral blood smears were compared between the DC-1, the DI-60 and manual microscopy to establish accuracy, within-run imprecision, clinical sensitivity and specificity. Moreover, the agreement between precharacterisation and postcharacterisation of red blood cell (RBC) morphological abnormalities was determined for the DC-1.

RESULTS

WBC preclassification and postclassification results of the DC-1 showed good correlation compared with DI-60 results and manual microscopy. In addition, the within-run SD of the DC-1 was below 1 for all five major WBC classes, indicating good reproducibility. Clinical sensitivity and specificity were, respectively, 96.7%/95.9% compared with the DI-60% and 96.6%/95.3% compared with manual microscopy. The overall agreement on RBC morphology between the precharacterisation and postcharacterisation results ranged from 49% (poikilocytosis) to 100% (hypochromasia, microcytosis and macrocytosis).

CONCLUSIONS

The DC-1 has proven to be an accurate digital cell imaging system for differential counting and morphological classification of WBCs and RBCs in peripheral blood smears. It is a compact and easily operated instrument that can offer low-volume and small satellite laboratories the possibilities of readily available blood cell analysis that can be stored and retrieved for consultation with remote locations.

Для цитирования: van der Vorm Lisa N, Hendriks Henriët A, Smits Simone M. Performance of the CellaVision DC-1 digital cell imaging analyser for differential counting and morphological classification of blood cells.. Journal of clinical pathology. 2021. DOI: 10.1136/jclinpath-2021-207863.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.