BACKGROUND

Journey II Bi-Cruciate-Stabilized knee system was designed to overcome the complications of Journey Bi-Cruciate-Stabilized, including ilio-tibial band inflammation and episodes of dislocation. The purpose of this study was to assess differences in knee kinematics between the first and second-generation design by means of video-fluoroscopy. Re-designed prosthesis in-vivo kinematics was analyzed during activities of daily living and results were eventually compared with those of the previous system, as reported in a previously published study. It was hypothesized that changes in components' design influences replaced knee's kinematic patterns.

METHODS

Sixteen patients (3 males, 13 females) implanted with the redesigned prosthesis were assessed by video-fluoroscopy during stair-climbing, chair-rising and leg-extension at 8 months of follow-up. Patterns of axial rotation and antero-posterior motion of the medial and lateral femoral condyles were obtained. Range of Motion and International Knee Society Score were recorded pre- and post-operatively. Student t-tests were applied to compare the mean of each interesting variables.

FINDINGS: The comparison of the kinematics of the two designs revealed similar patterns of axial rotation, with progressive femoral external rotation in flexion and reduced absolute values of displacement for the new system. Reduced posterior displacements of the medial and lateral condyles were observed in Journey II patients. In terms of absolute location, the lateral condyle in the redesigned prosthesis showed a more anterior position on the tibial-baseplate embedded coordinate system at maximal flexion.

INTERPRETATION: Design changes in the recently-introduced total knee system contributed to modify its in-vivo knee kinematics as demonstrated by video-fluoroscopy.

Для цитирования: Zambianchi Francesco, Fiacchi Francesco, Lombari Vincenzo, Venturelli Luca, Marcovigi Andrea, Giorgini Andrea, Catani Fabio. Changes in total knee arthroplasty design affect in-vivo kinematics in a redesigned total knee system: A fluoroscopy study.. Clinical biomechanics (Bristol, Avon). 2018. DOI: 10.1016/j.clinbiomech.2018.03.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.