Experimental Impact of Increasing Circuit Resistance in the Artificial Womb.

Cao Haiyan, Davey Marcus G, Young Kathleen, Tian Zhiyun, Kosaka Seitaro, Varela Maria F, Flake Alan W, Rychik Jack

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To determine the impact of controlled incremental changes in circuit blood flow resistance to the oxygenator on hemodynamics in the fetal lamb supported by the EXTrauterine Environment for Neonatal Development (EXTEND) system.

METHODS

A prospective study of oxygenator circuit clamping was conducted on 7 lambs on EXTEND. Echocardiography was performed at four levels of circuit flow (250, 225, 200 and 300 mL/kg/min). Hemodynamic parameters including cardiac output, velocities and pulsatility indexes (PI) were measured, and physiological parameters were documented.

RESULTS

As circuit resistance increases, combined cardiac output (CCO) declines significantly, with reduction most profoundly evident in the right ventricle with left ventricle flow preserved until reaching the lowest circuit flow level of 200 mL/kg/min. Umbilical artery (UA), umbilical vein (UV), and inferior vena cava (IVC) velocities decrease while middle cerebral artery (MCA) velocities increase. UA PI values change commensurately with changes in circuit resistance; however, MCA PI values did not change. Blood pressure and oxygen extraction elevate with increased circuit resistance, while heart rate and oxygen consumption remain unchanged.

CONCLUSIONS

Fetal sheep supported by the EXTEND system undergo hemodynamic and physiological changes in response to alterations in circuit flow, reflecting physiological adaptations to maintain circulatory homeostasis.

Для цитирования: Cao Haiyan, Davey Marcus G, Young Kathleen, Tian Zhiyun, Kosaka Seitaro, Varela Maria F, Flake Alan W, Rychik Jack. Experimental Impact of Increasing Circuit Resistance in the Artificial Womb.. Prenatal diagnosis. 2025. DOI: 10.1002/pd.6802.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.