DOI 10.1111/ped.13412

Медицинская библиотека · Pediatrics international : official journal of the Japan Pediatric Society · 2017

Loss of Surfacten® during bolus administration with a feeding catheter.

Fujioka Kazumichi, Kuroda Jumpei, Yamana Keiji, Iijima Kazumoto, Morioka Ichiro

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Surfactant replacement therapy is widely used for treating neonatal respiratory distress syndrome, but insufficient evidence is available on the use of Surfacten® (S-TA). This study investigated the inadvertent loss of S-TA during instillation via feeding catheters with different bore sizes.

METHODS

In this bench-based study, we measured the weight of syringes and tubes before and after surfactant treatment using a high-accuracy balance, and determined the amount of S-TA lost in tubes. We injected 120 mg of S-TA suspended in 4 or 3 mL into tubes followed with or without air boluses. Experiments were performed in triplicate. Percent weight loss of S-TA in each tube was calculated with or without air boluses.

RESULTS

Percent weight loss of S-TA was significantly higher in larger-bore tubes (P < 0.01, overall ANOVA), and was significantly lower after air bolus flushing in 3 Fr, 4 Fr, and 5 Fr tubes (P < 0.005, respectively). The 3 mL S-TA suspensions had a significantly higher percent loss than the 4 mL S-TA suspensions when using 4 Fr and 5 Fr tubes, and the 5 Fr closed system (P < 0.05, respectively).

CONCLUSIONS

Routine air bolus flushing effectively reduces S-TA loss in tubes. The 3 mL S-TA suspensions appear to be more susceptible to inadvertent S-TA loss during instillation. Therefore, caution is warranted for this procedure.

Для цитирования: Fujioka Kazumichi, Kuroda Jumpei, Yamana Keiji, Iijima Kazumoto, Morioka Ichiro. Loss of Surfacten® during bolus administration with a feeding catheter.. Pediatrics international : official journal of the Japan Pediatric Society. 2017. DOI: 10.1111/ped.13412.