MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ejmp.2022.02.003

Медицинская библиотека · Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2022

Dosimetric comparison between three-dimensional conformal radiotherapy followed by electron beam boost and volumetric modulated arc therapy using concomitant boost for the heart and cardiac segments in patients with left-sided breast cancer at risk for radiation-induced cardiac toxicity.

Murakami Yu, Murakami Yuki, Kamima Tatsuya, Oguchi Masahiko, Abo Natsumi, Takahashi Taro, Kaneko Masahiro, Nakano Masahiro, Matsubayashi Fumiyasu, Harada Arisa, Taguchi Senzo, Hashimoto Takeo, Yoshioka Yasuo

СтатьяПеревод в очереди

Аннотация

PURPOSE: We aimed to compare dosimetric parameters between three-dimensional conformal radiation therapy followed by electron beam boost (3D-CRT + EB) and volumetric modulated arc therapy using simultaneous integrated boost (SIB-VMAT) in left-sided breast cancer patients.

METHODS

This study included 57 patients with left-sided breast cancer who underwent SIB-VMAT. All patients had a computed tomography-based maximum heart distance of ≥ 1 cm and were prescribed a dose of 42.56 Gy/16 fractions to the planning target volume and a concomitant-boosted target dose of 53.2 Gy or 51.2 Gy. The 3D-CRT + EB plan was retrospectively created for the purpose of comparison using tangential fields with field-in-field technique followed by electron beam irradiation.

RESULTS

The doses to the clinical target volume significantly improved in the SIB-VMAT plans. All dosimetric parameters for the left anterior descending coronary artery (LAD) and LAD middle position (LAD mid) in the SIB-VMAT plans were significantly lower than those for 3D-CRT + EB plans (P < 0.01), while the doses to the heart, lung, contralateral breast and non-target tissue were decreased in the 3D-CRT + EB plans compared with those in the SIB-VMAT plans (e.g., 1.9 Gy vs. 2.9 Gy; P < 0.001 for the mean dose of heart).

CONCLUSIONS

SIB-VMAT significantly improved the dose to the target while reducing the doses to the LAD and LAD mid, whereas 3D-CRT + EB significantly decreased the doses to the heart and other organs at risk in patients with left-sided breast cancer at risk for radiation-induced coronary artery disease.

Для цитирования: Murakami Yu, Murakami Yuki, Kamima Tatsuya, Oguchi Masahiko, Abo Natsumi, Takahashi Taro, Kaneko Masahiro, Nakano Masahiro, Matsubayashi Fumiyasu, Harada Arisa, Taguchi Senzo, Hashimoto Takeo, Yoshioka Yasuo. Dosimetric comparison between three-dimensional conformal radiotherapy followed by electron beam boost and volumetric modulated arc therapy using concomitant boost for the heart and cardiac segments in patients with left-sided breast cancer at risk for radiation-induced cardiac toxicity.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2022. DOI: 10.1016/j.ejmp.2022.02.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.