DOI 10.1016/s0952-8180(00)00116-1
Uniport soft-tip, open-ended catheters versus multiport firm-tipped close-ended catheters for epidural labor analgesia: a quality assurance study.
Аннотация
STUDY
OBJECTIVE
To compare a multiport, firm-tipped, close-ended, epidural catheter (Portex catheter) with a uniport, open-ended, soft-tipped, wire-reinforced catheter (Arrow catheter) in obstetric patients.
STUDY
DESIGN
Prospective data collection for intradepartmental quality assurance.
SETTING
Obstetric unit in a tertiary care maternity hospital.
PATIENTS: 2612 patients requesting labor analgesia.
INTERVENTIONS: The Arrow catheter was used in 1,352 women and the Portex catheter in 1,260 women.
MEASUREMENTS AND MAIN
RESULTS
The incidence of unsatisfactory block were 3.3% and 4.4% with the Arrow and Portex catheters, respectively (p = 0.2). The catheter perforated the dura matter in 0.4% of cases with both catheters. The incidence of epidural venipuncture was 1.1% with the Arrow catheter and 5.7% with the Portex catheter (p = 0.0001). Paresthesias occurred in 6% of cases with the Arrow catheter and 11. 2 % of cases with the Portex catheter (p = 0.0001). Epidural catheter reinsertion was required in fewer patients in the Arrow group than in the Portex group (4.8% vs. 7.1%; p = 0.01).
CONCLUSIONS
In obstetric patients, the softer uniport Arrow catheter produces paresthesias and venipunctures less frequently than the firm multiport Portex catheter.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.