Effect of a comprehensive deep-learning model on the accuracy of chest x-ray interpretation by radiologists: a retrospective, multireader multicase study.

Seah Jarrel C Y, Tang Cyril H M, Buchlak Quinlan D, Holt Xavier G, Wardman Jeffrey B, Aimoldin Anuar, Esmaili Nazanin, Ahmad Hassan, Pham Hung, Lambert John F, Hachey Ben, Hogg Stephen J F, Johnston Benjamin P, Bennett Christine, Oakden-Rayner Luke, Brotchie Peter, Jones Catherine M

Оценочное исследованиеСтатьяВалидационное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Chest x-rays are widely used in clinical practice; however, interpretation can be hindered by human error and a lack of experienced thoracic radiologists. Deep learning has the potential to improve the accuracy of chest x-ray interpretation. We therefore aimed to assess the accuracy of radiologists with and without the assistance of a deep-learning model.

METHODS

In this retrospective study, a deep-learning model was trained on 821 681 images (284 649 patients) from five data sets from Australia, Europe, and the USA. 2568 enriched chest x-ray cases from adult patients (≥16 years) who had at least one frontal chest x-ray were included in the test dataset; cases were representative of inpatient, outpatient, and emergency settings. 20 radiologists reviewed cases with and without the assistance of the deep-learning model with a 3-month washout period. We assessed the change in accuracy of chest x-ray interpretation across 127 clinical findings when the deep-learning model was used as a decision support by calculating area under the receiver operating characteristic curve (AUC) for each radiologist with and without the deep-learning model. We also compared AUCs for the model alone with those of unassisted radiologists. If the lower bound of the adjusted 95% CI of the difference in AUC between the model and the unassisted radiologists was more than -0·05, the model was considered to be non-inferior for that finding. If the lower bound exceeded 0, the model was considered to be superior.

FINDINGS: Unassisted radiologists had a macroaveraged AUC of 0·713 (95% CI 0·645-0·785) across the 127 clinical findings, compared with 0·808 (0·763-0·839) when assisted by the model. The deep-learning model statistically significantly improved the classification accuracy of radiologists for 102 (80%) of 127 clinical findings, was statistically non-inferior for 19 (15%) findings, and no findings showed a decrease in accuracy when radiologists used the deep-learning model. Unassisted radiologists had a macroaveraged mean AUC of 0·713 (0·645-0·785) across all findings, compared with 0·957 (0·954-0·959) for the model alone. Model classification alone was significantly more accurate than unassisted radiologists for 117 (94%) of 124 clinical findings predicted by the model and was non-inferior to unassisted radiologists for all other clinical findings.

INTERPRETATION: This study shows the potential of a comprehensive deep-learning model to improve chest x-ray interpretation across a large breadth of clinical practice.

FUNDING: Annalise.ai.

Для цитирования: Seah Jarrel C Y, Tang Cyril H M, Buchlak Quinlan D, Holt Xavier G, Wardman Jeffrey B, Aimoldin Anuar, Esmaili Nazanin, Ahmad Hassan, Pham Hung, Lambert John F, Hachey Ben, Hogg Stephen J F, Johnston Benjamin P, Bennett Christine, Oakden-Rayner Luke, Brotchie Peter, Jones Catherine M. Effect of a comprehensive deep-learning model on the accuracy of chest x-ray interpretation by radiologists: a retrospective, multireader multicase study.. The Lancet. Digital health. 2021. DOI: 10.1016/s2589-7500(21)00106-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.