OBJECTIVE
To report outcomes of cranioplasty (CP) with a customized three-dimensional (3D) titanium mesh plate and identify the relationship between various clinical variables and complications after CP.
METHODS
Twenty patients who underwent surgery with customized 3D titanium mesh plates and had follow-up for >6 months from January 2015 to December 2017 were enrolled. Complications related to CP were retrospectively reviewed. Patients were divided into 2 groups (no complication group and complication group), and clinical variables related to complications were compared between the 2 groups. Correlation analysis was performed to identify the clinical variable significantly correlated with complications after CP with a customized 3D titanium mesh plate.
RESULTS
Of 20 patients, 11 developed complications, including extrusion, intracranial infection, severe operative site pain, and wound dehiscence. Among the clinical variables, only the presence of a ventriculoperitoneal shunt was significantly correlated with the overall complication rate (odds ratio = 18.66; 95% confidence interval, 1.56-222.92; P = 0.021).
CONCLUSIONS
A customized 3D titanium mesh plate is associated with a lower risk of intracranial infection, but the rate of postoperative complications, mainly cosmetic problems, such as forehead mesh extrusion and dehiscence, was high at 55%. A ventriculoperitoneal shunt was significantly correlated with the overall complication rate after CP with a customized 3D titanium mesh plate.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.