MedElevator Войти в исследования
← Поиск публикаций

DOI 10.5114/pjr/187934

Медицинская библиотека · Polish journal of radiology · 2024

The utility of virtual monochromatic dual-energy computed tomography (DECT) in meniscal imaging: a technical evaluation.

Adams Scott, van Heerden Jacques, Reis Jodie, Mondal Prosanta, Shepel Michael, Obaid Haron

СтатьяПеревод в очереди

Аннотация

PURPOSE: To determine the feasibility of meniscal imaging using virtual monochromatic images obtained through dualenergy computed tomography (DECT) technique, and to determine which keV levels optimise contrast resolution.

MATERIAL AND

METHODS

All DECT exams were performed on a Discovery CT750 HD system from GE Healthcare. Virtual monochromatic images were reconstructed at 40 keV, 73 keV, 106 keV, and 139 keV. Contrast resolution of the medial and lateral menisci using a 5-point Likert scale at each keV level was determined through a consensus agreement by 2 fellowship-trained musculoskeletal radiologists. Friedman's and Wilcoxon signed rank tests were used to compare visualisation scores across different keV levels.

RESULTS

Seventeen knee exams from 10 patients met criteria for inclusion in the study. All patients included in the study cohort were male. The median age of patients was 46 years (interquartile range, 35-53 years). Virtual monochromatic images at 40 keV demonstrated highest contrast resolution of the menisci, with a statistically significant difference between contrast resolution scores at 40 keV and 76 keV, Friedman test: p < 0.0001.

CONCLUSIONS

Meniscal imaging is feasible using DECT virtual monochromatic images at low keV levels. Improved contrast resolution at these specified KeV values could pave the way for further research in this field to determine its role in the future as an alternative option for assessment of the menisci in patients with contraindications to MRI or in the setting of a periarticular ferromagnetic foreign body obscuring the field of view.

Для цитирования: Adams Scott, van Heerden Jacques, Reis Jodie, Mondal Prosanta, Shepel Michael, Obaid Haron. The utility of virtual monochromatic dual-energy computed tomography (DECT) in meniscal imaging: a technical evaluation.. Polish journal of radiology. 2024. DOI: 10.5114/pjr/187934.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.