MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/eul172

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2007

Medium-term follow-up and modes of failure following epicardial pacemaker implantation in young children.

Bakhtiary Farhad, Dzemali Omer, Bastanier Christian K, Moritz Anton, Kleine Peter

СтатьяПеревод в очереди

Аннотация

AIMS: Young children suffering from congenital or post-operative AV-block require life-long pacemaker stimulation. Due to the anatomical prerequisites initially epicardial electrodes are implanted and the generator is placed in the upper abdominal wall. The following study investigated modes of failure leading to reoperation in this group of technically challenging patients.

METHODS AND

RESULTS

Between October 2000 and May 2005, a total of 21 infants (age 3 days to 5 years) underwent pacemaker implantation using a subxyphoidal incision for newborns (and a partial lower or complete median sternotomy for older children). Nine patients had previous cardiac surgery for complex congenital defects. The remaining 12 young children suffered from congenital AV-Block (CAVB). Twenty-one bipolar epicardial electrodes (Medtronic Capsure epi) were fixed to the right ventricle, 15 had additional implantation of a bipolar atrial lead. The pacemaker generator (Medtronic Kappa 701) was implanted into the right upper abdominal wall. Indications for revision were recorded. No mortality was observed; pacing and sensing parameters remained stable up to a 5-year follow-up. A total of four reoperations occurred. Three of the four revisions were caused by ventricular electrode fracture. At revision, two electrodes were broken at the crossing between the pericardial cavity and the abdominal wall, one bipolar lead at the Y-division into the two tip electrodes. One reoperation was due to a pacemaker recall. All revisions were performed without complications.

CONCLUSION

In our institute epicardial pacing in young children was associated with a satisfactory clinical outcome, but also a significant number of failures leading to reoperation, mainly due to electrode fracture caused by the muscular activity of this patient group. Reoperations were performed at a low risk.

Для цитирования: Bakhtiary Farhad, Dzemali Omer, Bastanier Christian K, Moritz Anton, Kleine Peter. Medium-term follow-up and modes of failure following epicardial pacemaker implantation in young children.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2007. DOI: 10.1093/europace/eul172.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.