DOI 10.1097/mbp.0000000000000799

Медицинская библиотека · Blood pressure monitoring · 2026

Accuracy of atrial fibrillation detection using Hanvon Korotkoff sound blood pressure monitor.

Wang Jian, Liu Yingjian, Chen Chongjie

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Atrial fibrillation is one of the most common cardiac arrhythmias. It is a condition that significantly increases the risk of stroke and cardiac complications, making its early diagnosis and intervention a crucial part of improving long-term patient prognosis. This study evaluates the accuracy of an atrial fibrillation detection algorithm in a Korotkoff sound blood pressure (BP) monitor.

METHODS

A total of 312 inpatients and outpatients were recruited for this study, with a Korotkoff sound BP monitor used to measure the BP and screen for atrial fibrillation. The study cohort included 148 patients with atrial fibrillation (mean age 73.3 ± 9.0 years) and 164 controls with sinus rhythm (mean age 44.7 ± 14.6 years). During BP measurement, a hand-held single-lead ECG device was used to simultaneously record ECGs for evaluating the accuracy of the atrial fibrillation detection algorithm.

RESULTS

The device accurately identified all 148 confirmed atrial fibrillation patients, reliably triggering 'AFib' alerts in every instance. Throughout the evaluation of 164 non-atrial fibrillation subjects, the device maintained perfect specificity with zero false positive events. These results demonstrate 100% sensitivity and 100% specificity for atrial fibrillation detection.

CONCLUSION

The Korotkoff sound BP monitor with atrial fibrillation detection algorithm exhibits excellent diagnostic accuracy. This device is of significant clinical value for early atrial fibrillation screening, serving as a convenient, efficient, and reliable tool for both home and primary medical institutions. Its capability for timely atrial fibrillation detection may contribute to reduced risks of stroke and other atrial fibrillation-related cardiovascular complications.

Для цитирования: Wang Jian, Liu Yingjian, Chen Chongjie. Accuracy of atrial fibrillation detection using Hanvon Korotkoff sound blood pressure monitor.. Blood pressure monitoring. 2026. DOI: 10.1097/mbp.0000000000000799.