Factors affecting electrogram sensing in an insertable cardiac monitor: Insights from surface electrocardiogram mapping analysis.

Pitman Bradley M, Zanker Amy, Lim Matthew, McLoughney Joshua, Spinelli Jade, Tarone Rachel, McInnes Kristie, Heath Kyle M, Gieve Mahsa, Evans Shaun, Young Glenn D, Roberts-Thomson Kurt C, Wong Christopher X, Sanders Prashanthan, Lau Dennis H

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Fidelity of electrogram sensing may reduce false alerts from an insertable cardiac monitor (ICM).

OBJECTIVE

The purpose of this study was to assess the effect of vector length, implant angle, and patient factors on electrogram sensing using surface electrocardiogram (ECG) mapping.

METHODS

Twelve separate precordial single-lead surface ECGs were acquired from 150 participants at 2 interelectrode distances (75 and 45 mm), at 3 vector angles (vertical, oblique, and horizontal), and in 2 postures (upright and supine). A subset of 50 patients also received a clinically indicated ICM implant in 1:1 ratio (Reveal LINQ [Medtronic, Minneapolis, MN]/BIOMONITOR III [Biotronik, Berlin, Germany]). All ECGs and ICM electrograms were analyzed by blinded investigators using DigitizeIt software (V2.3.3, Braunschweig, Germany). The P-wave visibility threshold was set at > 0.015 mV. Logistic regression was used to identify factors affecting P-wave amplitude.

RESULTS

A total of 1800 tracings from 150 participants (44.5% [n = 68] female; median age 59 [35-73] years) were assessed. The median P- and R-wave amplitudes were 45% and 53% larger with vector lengths of 75 and 45 mm, respectively (P < .001 for both). The oblique orientation yielded the best P- and R-wave amplitudes, while posture change did not affect P-wave amplitude. Mixed effects modeling found that visible P-waves occur more frequently with a vector length of 75 mm than with 45 mm (86% vs 75%, respectively; P < .0001). A longer vector length improved both P-wave amplitude and visibility in all body mass index categories. There was a moderate correlation of P- and R-wave amplitudes from the ICM electrograms to those from surface ECG recordings (intraclass correlation coefficient 0.74 and 0.80, respectively).

CONCLUSION

Longer vector length and oblique implant angle yielded the best electrogram sensing and are relevant considerations for ICM implantation procedures.

Для цитирования: Pitman Bradley M, Zanker Amy, Lim Matthew, McLoughney Joshua, Spinelli Jade, Tarone Rachel, McInnes Kristie, Heath Kyle M, Gieve Mahsa, Evans Shaun, Young Glenn D, Roberts-Thomson Kurt C, Wong Christopher X, Sanders Prashanthan, Lau Dennis H. Factors affecting electrogram sensing in an insertable cardiac monitor: Insights from surface electrocardiogram mapping analysis.. Heart rhythm. 2023. DOI: 10.1016/j.hrthm.2023.05.034.