DOI 10.1177/19714009231173099

Медицинская библиотека · The neuroradiology journal · 2023

Field strength difference in extent of artifacts induced by CERTAS Plus valves in patients with idiopathic normal pressure hydrocephalus.

Camerucci Emanuele, Elder Benjamin D, Shu Yunhong, Messina Steven A, Gunter Jeffrey L, Graff-Radford Jonathan, Jones David T, Botha Hugo, Cutsforth-Gregory Jeremy K, Jack Clifford R, Huston John, Cogswell Petrice M

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: : Post-shunt MRI is usually performed at 1.5T under the general assumption that shunt-related susceptibility artifacts would be greater at higher field strengths.

PURPOSE: The purpose is to show that imaging post-shunt idiopathic normal pressure hydrocephalus (iNPH) patients at 3T is feasible and with reduced artifacts as compared to 1.5T.

METHODS

We manually measured transverse dimensions of artifact at the levels of lateral ventricles, cerebral aqueduct, and cerebellar hemisphere. Areas/volumes of artifacts were calculated assuming an elliptic/ellipsoid shape. Relative extent of shunt-related artifact between field strengths was rated by 3 readers on a 5-point Likert scale. A Wilcoxon Signed Rank Test was used to compare artifact at 1.5T vs 3T for each sequence, with a significance level set at p < 0.05.

RESULTS

Artifact areas were calculated in 22 iNPH patients; artifacts were on average smaller at 3T vs 1.5T on MPRAGE, DWI, and GRE sequences. On T2 FLAIR and T2 FSE, artifacts at 3T were larger than 1.5T. On the qualitative analysis, artifact effects were less at 3T vs 1.5T on DWI, greater at 3T on T2 FSE, and had mixed results on GRE.

CONCLUSION

Our results indicate feasibility of post-shunt imaging with the CERTAS Plus valve at 3T based on shunt-related artifact that is less than or equal in extent to that on 1.5T on most standard clinical imaging sequences. Our findings, corroborated by the qualitative image review, suggest that dedicated clinical imaging sequences for devices may allow for reduction in artifact extent at both 1.5T and 3T.

Для цитирования: Camerucci Emanuele, Elder Benjamin D, Shu Yunhong, Messina Steven A, Gunter Jeffrey L, Graff-Radford Jonathan, Jones David T, Botha Hugo, Cutsforth-Gregory Jeremy K, Jack Clifford R, Huston John, Cogswell Petrice M. Field strength difference in extent of artifacts induced by CERTAS Plus valves in patients with idiopathic normal pressure hydrocephalus.. The neuroradiology journal. 2023. DOI: 10.1177/19714009231173099.

Предметные рубрики MeSH