DOI 10.1002/ccd.27852

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2018

Study of biodegradable occluder of atrial septal defect in a porcine model.

Li Bo-Ning, Xie Yu-Mei, Xie Zhao-Feng, Chen Xian-Miao, Zhang Gui, Zhang De-Yuan, Liu Xiang-Dong, Zhang Zhi-Wei

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the safety and feasibility of a modified poly(l-lactic acid) (PLLA) atrial septal defect (ASD) occluder.

METHODS

Forty-five piglets were divided into two groups: an experimental group (n = 27) and a control group (n = 18). The experimental group underwent percutaneous implantation of a modified PLLA ASD device while the control group underwent percutaneous implantation of a widely used metal ASD device. X-ray imaging, transthoracic echocardiography (TTE), electrocardiogram (ECG), histopathology and electron microscopic examination were performed at 7 days, 1, 3, 6, and 12 months after implantation.

RESULTS

Twenty-seven experimental piglets and 18 control piglets were all successfully implanted with modified biodegradable and metal ASD devices, respectively. While both devices exhibited very good occluding effects, the modified PLLA ASD devices were completely endothelialized at 3 months after implantation, and the endothelialization appeared to be more complete compared to the control group. Degradation of the PLLA devices was noted at 12 months follow-up with no loss of integrity at the atrial septum.

CONCLUSION

This animal model with implanting of the occluders was effective and not associated with complications. The modified PLLA ASD devices are more controllable and practical than our previous devices. The implanted devices demonstrated good endothelialization and degradability in short and moderate term follow-up. Long-term studies are now underway to further evaluate the biodegradability of this novel device.

Для цитирования: Li Bo-Ning, Xie Yu-Mei, Xie Zhao-Feng, Chen Xian-Miao, Zhang Gui, Zhang De-Yuan, Liu Xiang-Dong, Zhang Zhi-Wei. Study of biodegradable occluder of atrial septal defect in a porcine model.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2018. DOI: 10.1002/ccd.27852.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.