Validation of the mobil-O-Graph: 24 h-blood pressure measurement device.

Wei Wolfgang, Tölle Markus, Zidek Walter, van der Giet Markus

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Twenty-four-hour blood pressure measurement is of importance not only in the detection of hypertension but also in the detection of blood pressure changes in hypertensive and nonhypertensives over the day to identify, for example, nondipper hypertensives. This study describes the validation of the mobil-O-Graph according to the criteria of the British Hypertension Society (BHS).

METHODS

For each patient three readings obtained by the mobil-O-Graph were compared with auscultatory sphygmomanometric readings obtained by two trained clinicians. The sphygmomanometric reference measurements were alternated with the readings obtained by the device. Eighty-five patients (mean age 53.4+/-18.4 years) were recruited for the BHS protocol. Differences between blood pressure values of the test device and the mercury reading were calculated for each measurement.

RESULTS

In the BHS validation procedure the mean differences of the observer readings and the test device were -2.2+/-6.7 (systolic) and -0.6+/-5.6 mmHg (diastolic) for observer 1 and -2.2+/-7.3 mmHg (systolic) and-0.4+/-6.1 mmHg (diastolic) for observer 2. The device achieved grade A for systolic and diastolic blood pressure for both the observers 1 and 2 leading to a final grade A/A. According to the BHS protocol the measurements of the device have to be considered 'very accurate and with no error of clinical relevance'.

CONCLUSION

The device met the accuracy requirements of the BHS standard and can be recommended for clinical use.

Для цитирования: Wei Wolfgang, Tölle Markus, Zidek Walter, van der Giet Markus. Validation of the mobil-O-Graph: 24 h-blood pressure measurement device.. Blood pressure monitoring. 2010. DOI: 10.1097/mbp.0b013e328338892f.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.