MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11548-025-03434-w

Медицинская библиотека · International journal of computer assisted radiology and surgery · 2025

Semi-automatic segmentation of elongated interventional instruments for online calibration of C-arm imaging system.

Chabi Negar, Illanes Alfredo, Beuing Oliver, Behme Daniel, Preim Bernhard, Saalfeld Sylvia

СтатьяПеревод в очереди

Аннотация

PURPOSE: The C-arm biplane imaging system, designed for cerebral angiography, detects pathologies like aneurysms using dual rotating detectors for high-precision, real-time vascular imaging. However, accuracy can be affected by source-detector trajectory deviations caused by gravitational artifacts and mechanical instabilities. This study addresses calibration challenges and suggests leveraging interventional devices with radio-opaque markers to optimize C-arm geometry.

METHODS

We propose an online calibration method using image-specific features derived from interventional devices like guidewires and catheters (In the remainder of this paper, the term"catheter" will refer to both catheter and guidewire). The process begins with gantry-recorded data, refined through iterative nonlinear optimization. A machine learning approach detects and segments elongated devices by identifying candidates via thresholding on a weighted sum of curvature, derivative, and high-frequency indicators. An ensemble classifier segments these regions, followed by post-processing to remove false positives, integrating vessel maps, manual correction and identification markers. An interpolation step filling gaps along the catheter.

RESULTS

Among the optimized ensemble classifiers, the one trained on the first frames achieved the best performance, with a specificity of 99.43% and precision of 86.41%. The calibration method was evaluated on three clinical datasets and four phantom angiogram pairs, reducing the mean backprojection error from 4.11 ± 2.61 to 0.15 ± 0.01 mm. Additionally, 3D accuracy analysis showed an average root mean square error of 3.47% relative to the true marker distance.

CONCLUSIONS

This study explores using interventional tools with radio-opaque markers for C-arm self-calibration. The proposed method significantly reduces 2D backprojection error and 3D RMSE, enabling accurate 3D vascular reconstruction.

Для цитирования: Chabi Negar, Illanes Alfredo, Beuing Oliver, Behme Daniel, Preim Bernhard, Saalfeld Sylvia. Semi-automatic segmentation of elongated interventional instruments for online calibration of C-arm imaging system.. International journal of computer assisted radiology and surgery. 2025. DOI: 10.1007/s11548-025-03434-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.