Wear investigation of implant-supported upper removable prothesis with electroplated gold or PEKK secondary crowns.
Аннотация
OBJECTIVES
The aim of the present study was to investigate, whether polyetherketoneketone (PEKK) secondary crowns could be considered as alternative to gold standard in terms of their physical properties and manufacturing costs.
METHODS
An upper jaw model with six implants was used. Frameworks with either 6 PEKK- or 6 electroplated secondary crowns were cemented in a wear simulator. A total of 20 specimens (10 PEKK, 10 gold) run 10,000 cycles in the wear simulator with a lubricant. Additionally, 10,000 cycles in the thermocycling baths with 5 °C and 55 °C have gone through, before running extra 10,000 cycles in the wear simulator again. Finally, the abutments were analysed for signs of wear under the electron microscope.
RESULTS
The mean pulling out force value for PEKK was 21 N. For the electroplated gold secondary crowns an average of 19 N was measured. Multiple fluctuations were observed in the gold series of tests. After 20,000 cycles in the wear simulator and 10,000 cycles in the thermocycling machine, there were no major losses to be measured in terms of wear for both materials. In the microscopic analysis of the abutments, traces of wear could be seen in pull-out direction, mainly in the gold samples.
CONCLUSIONS
PEKK secondary crowns have lower costs, more stable retention force values and are easier to produce than the gold standard. On average, the pull-out force values were 11 N higher than recommended.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.