MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s12903-025-05486-5

Медицинская библиотека · BMC oral health · 2025

Biomechanical behavior of titanium, cobalt-chromium, zirconia, and PEEK frameworks in implant-supported prostheses: a dynamic finite element analysis.

Sahin Hazir Dilara, Sozen Yanik Irem, Guncu M Baris, Canay R Senay

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The mechanical properties of framework materials significantly influence stress distribution and the long-term success of implant-supported prostheses. Although titanium, cobalt-chromium, zirconia, and polyether ether ketone (PEEK) are widely used, their biomechanical performance under dynamic loading conditions remains insufficiently investigated. This study aimed to evaluate the biomechanical behavior of four framework materials with different Young's modulus using dynamic finite element stress analysis.

METHODS

A 3D edentulous maxillary model was extracted from a computer tomography (CT) database. Bone level implants with conical connection designs were placed in the anterior (canine) and posterior (first molar) areas. Anterior implants were parallel, yet posterior implants were inclined distally by 30 degrees. According to the framework material, four groups were formed: cobalt-chromium (Co-Cr), zirconia (Zr), titanium (Ti), and polyether ether ketone (PEEK). For each framework material, twelve separate models of analysis were created by applying force in three different orientations. Dynamic forces were employed to replicate the chewing process. Principal and von Mises stresses were measured and evaluated.

RESULTS

The PEEK framework exhibited the highest maximum von Mises stress values (372.55 MPa) on the abutment and the highest maximum principle stress values (59.27 MPa) in the cortical bone. The Co-Cr framework had the lowest minimum principle stress (3.98 MPa) in the trabecular bone. The displacements of the Co-Cr, Zr, Ti, and PEEK frameworks were 0.15 mm, 0.15 mm, 0.17 mm, and 0.35 mm, respectively.

CONCLUSION

Frameworks having a greater Young's modulus are less susceptible to deformation.

Для цитирования: Sahin Hazir Dilara, Sozen Yanik Irem, Guncu M Baris, Canay R Senay. Biomechanical behavior of titanium, cobalt-chromium, zirconia, and PEEK frameworks in implant-supported prostheses: a dynamic finite element analysis.. BMC oral health. 2025. DOI: 10.1186/s12903-025-05486-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.