MedElevator Войти в исследования
← Поиск публикаций

DOI 10.7759/cureus.108925

Медицинская библиотека · Cureus · 2026

Fluorescein Interferes With Hemoglobin Measurement During Intra-operative Point-of-Care Blood Gas Analysis: A Clinical Observation.

Kojima Yuki, Moriwaki Takuya, Kitamura Haruka, Nakano Kazuhiko, Hirabayashi Kazuya

Редакционная статьяПеревод в очереди

Аннотация

We highlight a clinically important but under-recognized interference in hemoglobin measurement caused by intravenous fluorescein during neurosurgical procedures. A 70-year-old woman undergoing brain tumor resection showed normal initial blood gas results, but repeated point-of-care co-oximetry failed to measure hemoglobin after fluorescein administration. An alternative method enabled successful measurement of hemoglobin and transfusion guidance. Similar measurement failures were recorded previously, with failure in 7 of 12 cases. The fact that analyzers using different absorbance wavelengths were less affected suggests that spectral overlap between fluorescein and co-oximetry wavelengths may underlie the interference. A volunteer study further demonstrated that measurement failure occurred shortly after fluorescein administration but resolved within approximately 1 h. Clinicians should be aware of this limitation and consider alternative or confirmatory methods, such as central laboratory testing, when accurate hemoglobin assessment is critical.

Для цитирования: Kojima Yuki, Moriwaki Takuya, Kitamura Haruka, Nakano Kazuhiko, Hirabayashi Kazuya. Fluorescein Interferes With Hemoglobin Measurement During Intra-operative Point-of-Care Blood Gas Analysis: A Clinical Observation.. Cureus. 2026. DOI: 10.7759/cureus.108925.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.