DOI 10.1093/jat/22.2.105

Медицинская библиотека · Journal of analytical toxicology · 1998

A barbiturate screening assay for the Abbott AxSYM analyzer.

Adamczyk M, Douglas J, Grote J, Harrington C A

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

A fluorescence polarization immunoassay for barbiturates on the Abbott AxSYM analyzer is described. The assay displayed dilution linearity up to 1200 ng/mL; coefficients of variation varied from 5.96 to 8.61%; recovery varied from 94.9 to 105.3%; and sensitivity was less than 40 ng/mL. Good correlation between the standard six- and factory two-point calibration methods was observed. The Immunoassay demonstrated good cross-reactivity to several commonly prescribed barbiturates; low cross-reactivity with structurally similar compounds; low interference from endogenous substances, dyes, preservatives, and several commonly available adulterants; and good correlation with the TDx Barbiturate Urine assay.

Для цитирования: Adamczyk M, Douglas J, Grote J, Harrington C A. A barbiturate screening assay for the Abbott AxSYM analyzer.. Journal of analytical toxicology. 1998. DOI: 10.1093/jat/22.2.105.