DOI 10.1016/j.talanta.2008.06.020

Медицинская библиотека · Talanta · 2008

A novel biamperometric biosensor for urinary oxalate determination using flow-injection analysis.

Milardović Stjepan, Kereković Irena, Nodilo Marijana

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

A biosensor for determination of oxalate concentration in urine has been developed by immobilisation of oxalate oxidase and peroxidase on the surface of an interdigitated gold electrode. Enzyme immobilisation was performed using BSA and glutaraldehyde. Biamperometric measurements were made in flow conditions both in aqueous oxalate solutions (tested concentration range between 50 microM and 10 mM) and in real urine samples (tested measuring range between 5 and 100 microM). Optimal working conditions were examined for flow-injection analysis, and good correlation was achieved between added oxalate quantity and the one measured by biosensor in urine matrix (R(2)=0.9983). The influence of some interferences (ascorbic acid, uric acid, paracetamol, acetylsalicylic acid) was also studied using biamperometric measurement mode.

Для цитирования: Milardović Stjepan, Kereković Irena, Nodilo Marijana. A novel biamperometric biosensor for urinary oxalate determination using flow-injection analysis.. Talanta. 2008. DOI: 10.1016/j.talanta.2008.06.020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.