DOI 10.1016/j.jacep.2022.12.022

Медицинская библиотека · JACC. Clinical electrophysiology · 2023

Comparison of Pacing Performance and Clinical Outcomes Between Left Bundle Branch and His Bundle Pacing.

Tan Eugene S J, Soh Rodney, Boey Elaine, Lee Jie-Ying, de Leon Jhobeleen, Chan Siew-Pang, Gan Hiong-Hiong, Seow Swee-Chong, Kojodjojo Pipin

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Left bundle branch (LBBP) and His-bundle pacing (HBP) provide physiological ventricular activation.

OBJECTIVES

This study investigated differences in feasibility, device performance, and clinical outcomes between LBBP and HBP.

METHODS

Consecutive patients with LBBP and HBP from 2018 to 2021 in 2 centers were prospectively studied. The primary endpoint was optimal device performance during follow-up, defined as the presence of pacing thresholds <2.5 V, R-wave amplitude ≥5 V, and absence of conduction system pacing (CSP)-related complications. The secondary endpoint was the composite of heart failure hospitalizations or all-cause mortality.

RESULTS

Among 338 patients, 282 underwent successful CSP (119 HBP, 163 LBBP). Success rates, CSP-related complications, and need for reoperations did not differ between LBBP and HBP (P > 0.05). Pacing thresholds were lower, whereas R-wave amplitudes and lead impedance were higher in LBBP (P < 0.05). The primary endpoint was more frequent in LBBP than HBP (79% vs 34%; P < 0.001), with LBBP independently associated with 9-fold increased odds of optimal device performance (adjusted OR: 9.31; 95% CI: 5.14-16.86). LBBP was less likely to have increased pacing thresholds by >1 V (1% vs 19% HBP, P < 0.001). The secondary outcome was less frequent in LBBP than HBP (9% vs 24%, P = 0.001), with LBBP trending towards higher event-free survival (HR: 0.62; 95% CI: 0.31-1.23). The secondary outcome was independent of pacing burden or pacing indication.

CONCLUSIONS

Despite similar feasibility and safety profiles, LBBP confers additional benefits in pacing performance and reliability, shows trends towards improved survival compared to HBP, and should be the preferred first-line CSP modality of choice.

Для цитирования: Tan Eugene S J, Soh Rodney, Boey Elaine, Lee Jie-Ying, de Leon Jhobeleen, Chan Siew-Pang, Gan Hiong-Hiong, Seow Swee-Chong, Kojodjojo Pipin. Comparison of Pacing Performance and Clinical Outcomes Between Left Bundle Branch and His Bundle Pacing.. JACC. Clinical electrophysiology. 2023. DOI: 10.1016/j.jacep.2022.12.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.