DOI 10.1007/s40477-026-01212-1
Automated breast ultrasound (ABUS) meets artificial intelligence: a new accurate classification tool for BIRADS 3-4 breast lesions.
Аннотация
PURPOSE: Breast ultrasound (US) has often interpretation challenges (BIRADS 3-BIRADS 4 lesions), leading to a high demand for core needle biopsies (CNB). In this context, a computer-aided diagnostic (CAD) tool based on a machine-learning artificial intelligence, applied to images obtained with an automated breast ultrasound system (ABUS), could potentially provide more accurate diagnoses by instantly and automatically assessing the US lesion features and their level of suspicion. The aim of this study is to evaluate the diagnostic accuracy of a CAD applied to ABUS images for BIRADS 3-4 lesions.
MATERIALS AND
METHODS
Fifty-four patients underwent ABUS, performed with three scans per side, reconstructing images in the axial, coronal, and sagittal planes. Two expert radiologists draw a region of interest (ROI) on the lesion to be assessed by CAD tool, reaching a consensus regarding the images. Only breast lesions classified as BIRADS 3-4 were included, for a total of 68 (32 BIRADS 3, 36 BIRADS 4). CAD results were compared with the histological findings from CNBs in all cases, calculating sensitivity (SE), specificity (SP), accuracy (AC), positive predictive value (PPV), and negative predictive value (NPV).
RESULTS
Histology identified 40 positive lesions and 28 negative lesions. The CAD classified 38 suspicious lesions, with 6 false positives, 22 true negatives, and 2 false negatives. The values for SE, SP, AC, PPV, and NPV were 95, 79, 88, 86, and 92%, respectively.
CONCLUSIONS
The proposed CAD system is accurate in characterizing BIRADS 3-4 breast lesions diagnosed by ABUS and, therefore, could be applied in the clinical practice in order to assist radiologists in evaluating US images and reducing the number of the required CNBs.