MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.echo.2017.10.004

Медицинская библиотека · Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography · 2017

Quantification of Aortic Valve Regurgitation by Pulsed Doppler Examination of the Left Subclavian Artery Velocity Contour: A Validation Study with Cardiovascular Magnetic Resonance Imaging.

Spampinato Ricardo A, Jahnke Cosima, Paetsch Ingo, Hilbert Sebastian, Busch Franziska, Schloma Valerie, Dmitrieva Yaroslava, Bonamigo Thome Fernanda, Löbe Susanne, Strotdrees Elfriede, Hindricks Gerhard, Mohr Friedrich-Wilhelm, Borger Michael A

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Reflux of the aortic regurgitation (AR) causes an increased diastolic reverse flow in the aorta and its branching vessels. We aimed to evaluate the feasibility and accuracy of Doppler measurements in the left subclavian artery (LSA) for quantification of AR in a cardiovascular magnetic resonance imaging (CMR) validation study.

METHODS

Systolic and diastolic flow profiles of the LSA (subclavicular approach) were evaluated prospectively by use of pulsed wave Doppler in 59 patients (55.5 ± 15 years; 44 men), 47 with a wide spectrum of AR and 12 as control group. Using CMR phase-contrast sequences (performed 1 cm above the aortic valve), the AR was divided into three groups: mild, regurgitant fraction (RF) < 20% (n = 17); moderate, RF 20%-40% (n = 10); and severe, RF > 40% (n = 20). The LSA Doppler-derived RF was calculated as the ratio between diastolic and systolic velocity-time integrals (VTI).

RESULTS

Quality LSA Doppler signal could be obtained in all cases. Patients with CMR severe AR had higher values of LSA Doppler-derived RF (51% ± 9% vs 36% ± 11% vs 16% ± 8%; P < .0001). LSA Doppler showed a good correlation with CMR, with a sensitivity of 95%, specificity of 89%, and diagnostic accuracy for severe AR of 91.5%. Finally, Bland-Altman plots showed agreement in the group with moderate to severe AR (mean bias = -2.2% ± 8%, 95% CI, -17.7 to 13.3; P = .145) but differed in mild AR.

CONCLUSIONS

Measurements of the RF for quantification of AR using LSA Doppler are comparable to those of CMR, highlighting the potential role of LSA Doppler as an adjunctive technique to assess the severity of AR.

Для цитирования: Spampinato Ricardo A, Jahnke Cosima, Paetsch Ingo, Hilbert Sebastian, Busch Franziska, Schloma Valerie, Dmitrieva Yaroslava, Bonamigo Thome Fernanda, Löbe Susanne, Strotdrees Elfriede, Hindricks Gerhard, Mohr Friedrich-Wilhelm, Borger Michael A. Quantification of Aortic Valve Regurgitation by Pulsed Doppler Examination of the Left Subclavian Artery Velocity Contour: A Validation Study with Cardiovascular Magnetic Resonance Imaging.. Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography. 2017. DOI: 10.1016/j.echo.2017.10.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.