Аннотация

PURPOSE: To demonstrate the viability of a coordinate-measuring machine (CMM) for the geometric analysis of 3D printed surgical templates.

METHODS

The template was designed and modified by adding 18 cylindrical landmarks for CMM test and then classified into five groups according to the slicing software and resins (opaque and transparent): Streamflow-O, Streamflow-T, Shapeware-T, Rayware-T and Polydevs-T (N = 3). Three standing times (0 w, 1 w, and 2 w) were included to observe possible deformation. All the measurements were performed automatically by the CMM through a preset program. The Euclidian distance (dxyz) was regarded as the representation of global dimension accuracy, and displacements in the x-, y-, and z-axes were also calculated.

RESULTS

The average dxyz values of Streamflow-O, Streamflow-T, Shapeware-T, Rayware-T and Polydev-T are 32.6 μm, 31.3 μm, 56.4 μm, 96.4 μm, and 55.3 μm, respectively. Deviations were mainly induced by the upward bending of the free end region (positive direction of the z-axis). Different resins did not have a significant influence on the dimensional accuracy. Moreover, deformation appeared to be negligible after 2 weeks of storage, and the z-axis displacements were only approximately 30 μm at week 1 and 10 μm at week 2.

CONCLUSIONS

The deviations of the DLP-printed template are induced mainly by z-axis displacements and are determined by the processing accuracy. After 2 weeks, the dimensional stabilities of these templates are reliable, which is encouraging for clinicians. Moreover, the CMM is preliminarily demonstrated to be a feasible tool for achieving automated geometric analysis of surgical templates.

Для цитирования: He Lixing, Qin Bowen, Zhu Rongrong, Liu Yunxian, Xu Boya, Li Zhe, Du Liangzhi. Achieving automated and high-precision in situ analysis of the dimensional accuracy and dynamic deformation of 3D-printed surgical templates: an in vitro study.. International journal of implant dentistry. 2024. DOI: 10.1186/s40729-024-00561-y.