MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.carrev.2025.09.001

Медицинская библиотека · Cardiovascular revascularization medicine : including molecular interventions · 2025

A pharmacokinetic comparison between three drug delivery devices in porcine coronary arteries.

Tanaka Takamasa, Kawakami Rika, Shiraki Tatsuya, Nakayama Takafumi, Fujiyoshi Kazuhiro, Hamana Tomoyo, Adachi Yusuke, Sakamoto Atsushi, Grogan Alyssa, Virmani Renu, Finn Aloke V

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Catheter-based drug delivery systems have evolved in percutaneous coronary interventions (PCI). Drug-coated balloons (DCBs) deliver agents to obstructive atherosclerotic lesions without permanent implants, but their pharmacokinetic profiles vary and may affect efficacy and safety.

METHOD: Twenty-seven Yucatan Miniature Swine underwent treatment with SEL-SEB (SELUTION SLR™), MT-SCB (MagicTouch™), or EES (XIENCE™) in the right coronary or left circumflex arteries. Drug levels were measured in treated arteries and downstream myocardium at 7, 60, and 90 days. MT-SCB samples were extracted with ZnSO₄/methanol, whereas SEL-SEB samples underwent two-step extraction (ZnSO₄/methanol followed by acetonitrile/acetone).

RESULT: A total of 54 arteries underwent pharmacokinetic evaluation (n = 6 treated arteries per time point) from 27 swine. At 7 days, the mean of arterial drug levels was 1767.5 ng/g for SEL-SEB, 1136.5 ng/g for MT-SCB, and 1057.4 ng/g for EES (p = 0.94). By 60 days, SEL-SEB and EES maintained high levels (1146.3 and 837.3 ng/g), while MT-SCB fell sharply to 5.8 ng/g (p < 0.05). At 90 days, SEL-SEB and EES remained stable (360.7 ng/g and 916.5 ng/g, respectively), although MT-SCB was negligible, 2.0 ng/g (p < 0.05). Drug levels in the downstream myocardium at 7 days were lowest with EES (0.1 ng/g), higher with SEL-SEB (11.3 ng/g), and highest with MT-SCB (281.5 ng/g) (p < 0.05).

CONCLUSIONS

SEL-SEB shows sustained arterial drug retention, more closely resembling the gold-standard EES. This data suggests that SEL-SEB should produce durable outcomes without the need for a permanent implant.

Для цитирования: Tanaka Takamasa, Kawakami Rika, Shiraki Tatsuya, Nakayama Takafumi, Fujiyoshi Kazuhiro, Hamana Tomoyo, Adachi Yusuke, Sakamoto Atsushi, Grogan Alyssa, Virmani Renu, Finn Aloke V. A pharmacokinetic comparison between three drug delivery devices in porcine coronary arteries.. Cardiovascular revascularization medicine : including molecular interventions. 2025. DOI: 10.1016/j.carrev.2025.09.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.