Propensity score analysis of stented versus rapid deployment aortic bioprostheses in patients with small aortic annulus.

Cibin Giorgia, D'Onofrio Augusto, Lombardi Valentina, Bergonzoni Emma, Lorenzoni Giulia, Gastino Elisa, Evangelista Giuseppe, Italiano Enrico Giuseppe, Cao Irene, Gregori Dario, Tessari Chiara, Gerosa Gino

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

Haemodynamic studies have demonstrated the excellent performance of rapid deployment (RD) valves. This retrospective single-centre study aimed to compare early and medium-term outcomes of RD bioprostheses versus conventional stented valves in patients with small aortic annuli.

METHODS

We included patients who underwent isolated or combined surgical aortic valve replacement (SAVR) with Magna Ease (ME) and Intuity (Edwards Lifesciences, Irvine, CA) sizes 19 and 21 at our institution between June 2016 and March 2022. Follow-up was conducted through scheduled visits and echocardiograms at the study site, or via telephonic interviews with patients and/or referring cardiologists. A propensity score weighting analysis was performed to account for baseline differences between the 2 cohorts.

RESULTS

A total of 666 consecutive patients underwent SAVR with the 2 devices. ME was implanted in 367 patients (55.1%) and Intuity in 299 (44.9%). ME size 19 or 21 was used in 105 patients (35.1%), and Intuity size 19 or 21 in 115 patients (31.3%). Our study population comprised 220 patients. There were no significant differences in postoperative complications. Intuity demonstrated significantly lower gradients overall (mean gradients: 12 mmHg vs 16 mmHg, P < 0.001) and for size 21 (mean gradients: 12 mmHg vs 15 mmHg, P < 0.001). Mid-term survival and rehospitalization rates were similar between the 2 devices (5-year rehospitalization rate: 17% ME vs 20.9% Intuity, P = 0.57; 5-year survival: 81.9% ME vs 88% Intuity, P = 0.761).

CONCLUSIONS

In patients with small aortic annuli, RD bioprostheses provide superior haemodynamic outcomes compared to conventional stented valves. However, perioperative outcomes, mid-term survival, and rehospitalization rates are similar between the 2 devices.

Для цитирования: Cibin Giorgia, D'Onofrio Augusto, Lombardi Valentina, Bergonzoni Emma, Lorenzoni Giulia, Gastino Elisa, Evangelista Giuseppe, Italiano Enrico Giuseppe, Cao Irene, Gregori Dario, Tessari Chiara, Gerosa Gino. Propensity score analysis of stented versus rapid deployment aortic bioprostheses in patients with small aortic annulus.. Interdisciplinary cardiovascular and thoracic surgery. 2025. DOI: 10.1093/icvts/ivaf241.