DOI 10.17235/reed.2019.5924/2018
A novel rescue maneuver for a distal dislodged lumen-apposing metal stent (LAMS): "LAMS-in-LAMS" technique.
Аннотация
INTRODUCTION
the appearance of the lumen-apposing metal stent (LAMS) has meant an authentic revolution. To date, the results are promising but it is necessary to note the technical incidents and LAMS-related complications.
CASE REPORT: an EUS-transmural guided drainage using a HotAXIOS was planned for a 36-year-old man with oral intolerance due to a voluminous walled-off necrosis. The distal flange was left in the collection, but a total distal malposition occurred during the proximal flange delivery, despite correct apposition with visualization of the black mark. A rescue technique was performed inserting a second LAMS over-the-guidewire salvaging the initial failed transmural drainage.
DISCUSSION: This case is a reminder that in similar scenarios, extreme tension of the echoendoscope can cause a malfunction of the AXIOS stent delivery system, and lead to a total distal malposition. This "LAMS-in-LAMS" technique is feasible, effective, and a very helpful rescue technique in cases of dislodged LAMS.