DOI 10.1227/01.neu.0000328392.98602.5a

Медицинская библиотека · Neurosurgery · 2009

Assessment of zero drift in the Codman intracranial pressure monitor: a study from 2 neurointensive care units.

Al-Tamimi Yahia Z, Helmy Adel, Bavetta Seb, Price Stephen J

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Intraparenchymal monitoring devices play an important role in the daily management of head injury and other critically ill neurosurgical patients. Although zero drift data exist for the Camino system (Camino Laboratories, San Diego, CA), only in vitro data exist for the Codman system (Codman and Shurtleff, Inc., Raynham, MA). The aim of this study was to assess the extent of zero drift for the Codman intracranial pressure (ICP) monitor in patients being monitored in 2 neurointensive care units.

METHODS

This was a prospective study conducted at 2 neurointensive care units. Eighty-eight patients who required ICP monitoring and who presented to the 2 neurosurgical departments, Center 1 (n = 48) and Center 2 (n = 40), were recruited for participation. The duration of ICP monitoring was noted, as was the resultant pressure reading in normal saline on removing the ICP monitor (zero drift).

RESULTS

The median absolute zero drift for the group was 2.0 mm Hg (interquartile range, 1-3 mm Hg). The median time in situ was 108 hours (interquartile range, 69-201 hours). There was a positive correlation between the drift and time of the probe spent in situ (Spearman's correlation coefficient = 0.342; P = 0.001). Of the readings, 20 and 2% showed a drift greater than 5 and 10 mm Hg in magnitude, respectively.

CONCLUSION

These data demonstrate that a small amount of zero drift exists in ICP monitors and that this drift increases with time. The wide range in the data demonstrates that some drift readings are quite excessive. This reinforces the school of thought that, although ICP readings contribute significantly to the management of neurosurgical patients, they should be interpreted carefully and in conjunction with clinical and radiological assessment of patients.

Для цитирования: Al-Tamimi Yahia Z, Helmy Adel, Bavetta Seb, Price Stephen J. Assessment of zero drift in the Codman intracranial pressure monitor: a study from 2 neurointensive care units.. Neurosurgery. 2009. DOI: 10.1227/01.neu.0000328392.98602.5a.