Аннотация

OBJECTIVES

In the context of postcardiotomy cardiogenic shock (PCCS) following valve replacement surgery, it may be necessary to implant a peripheral veno-arterial extracorporeal membrane oxygenation (pVA-ECMO). This procedure, however, carries a risk of prosthetic valve thrombosis. The aim of this retrospective study was to describe the incidence and outcomes of prosthetic valve thrombosis after VA-ECMO support for PCCS and to report the associated risk factors.

METHODS

All consecutive adult patients who received pVA-ECMO for PCCS following a valve replacement procedure between January 2015 and October 2019 in our institution were included in this retrospective study. Outcome variables were prosthetic valve thrombosis, 30-day and hospital survival, pVA-ECMO-associated adverse events and surgery-related adverse events.

RESULTS

During the 4-year study period, 549 patients received pVA-ECMO for PCCS. Among them, 152 had undergone a valve replacement procedure and 9 of these developed prosthetic valve thrombosis. The incidence of valve thrombosis at 30 days was 7.5 ± 2%. The cumulative incidence of prosthetic valve thrombosis was significantly lower with pVA-ECMO + intra-aortic balloon pump versus VA-ECMO alone (1.4 ± 1.4% vs 13.7 ± 4.7%, P = 0.021, respectively). Intra-aortic balloon pump use associated with pVA-ECMO (versus pVA-ECMO alone) was an independent protective factor against hospital death [odds ratio = 0.180 (0.068-0.478), P = 0.001].

CONCLUSIONS

After PCCS following valve replacement surgery, peripheral femoro-femoral VA-ECMO is associated with a low risk of acute valve thrombosis especially when associated with an intra-aortic balloon pump.

Для цитирования: Danial Pichoy, Zamorano Claudio, Carillion Aude, Barreda Eleodoro, Laali Mojgan, Demondion Pierre, D'Alessandro Cosimo, Bouglé Adrien, Pineton de Chambrun Marc, Combes Alain, Leprince Pascal, Lebreton Guillaume. Incidence and outcomes of prosthetic valve thrombosis during peripheral extracorporeal membrane oxygenation.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2024. DOI: 10.1093/ejcts/ezae321.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.