Fox Henrik, Bitter Thomas, Horstkotte Dieter, Oldenburg Olaf, Gutleben Klaus-Jürgen
Аннотация
BACKGROUND: Sleep-disordered breathing (SDB) and Cheyne-Stokes respiration (CSR) are associated with shorter survival in patients with heart failure. A novel treatment method for this patient group is unilateral phrenic nerve stimulation by the remedē® system (Respicardia Inc., Minnetonka, MN, USA), a transvenously implantable neurostimulation device, which has recently been studied in a large randomized, controlled trial. Previous literature has shown efficacy and safety of the treatment with this first-generation device, but hardly any data are available on long-term clinical parameters, the remedē® device's battery lifetime, device exchangeability, lead position stability, surgical accessibility, and manageability.
METHODS: We performed remedē® device replacements in consecutive patients for battery depletion, and documented clinical parameters, longevity, operation procedure, complications, and difficulties.
RESULTS: All patients were on neurostimulation treatment by phrenic nerve neurostimulation when device replacement became necessary. Apnea-hypopnea index (from 45 ± 4/h to 9 ± 4/h), oxygen-desaturation index (from 35 ± 7/h to 7 ± 6/h), and time spent with oxygen saturation of <90% (T < 90% from 5 ± 7% to 0 ± 0%) were improved and improvements remained constant throughout the 4-year follow-up. Mean battery life was 4.2 ± 0.2 years and mean replacement procedure time was 25 ± 5.1 minutes. Apart from conventional X-ray documentation of stable lead positions in a long-term setting, no radiation or contrast dye usage was needed and no major complications occurred. In addition, clinical exercise capacity and sleepiness symptoms improved.
CONCLUSIONS: Novel remedē® device shows sustained therapy efficacy and safety in terms of stable lead positions over 4 years. Long-term phrenic nerve neurostimulation therapy for central SDB/CSR appears feasible in a clinical routine setting.
METHODS: We performed remedē® device replacements in consecutive patients for battery depletion, and documented clinical parameters, longevity, operation procedure, complications, and difficulties.
RESULTS: All patients were on neurostimulation treatment by phrenic nerve neurostimulation when device replacement became necessary. Apnea-hypopnea index (from 45 ± 4/h to 9 ± 4/h), oxygen-desaturation index (from 35 ± 7/h to 7 ± 6/h), and time spent with oxygen saturation of <90% (T < 90% from 5 ± 7% to 0 ± 0%) were improved and improvements remained constant throughout the 4-year follow-up. Mean battery life was 4.2 ± 0.2 years and mean replacement procedure time was 25 ± 5.1 minutes. Apart from conventional X-ray documentation of stable lead positions in a long-term setting, no radiation or contrast dye usage was needed and no major complications occurred. In addition, clinical exercise capacity and sleepiness symptoms improved.
CONCLUSIONS: Novel remedē® device shows sustained therapy efficacy and safety in terms of stable lead positions over 4 years. Long-term phrenic nerve neurostimulation therapy for central SDB/CSR appears feasible in a clinical routine setting.