Performance of the transoesophageal echocardiography probe as an oesophageal temperature monitor in patients undergoing cardiac surgery with cardiopulmonary bypass: a prospective observational study.
Аннотация
OBJECTIVES
Core temperature monitoring is critical during cardiopulmonary bypass (CPB). In this prospective observational study, we investigated the performance of the transoesophageal echocardiography (TOE) probe for core (oesophageal) temperature monitoring during CPB.
METHODS
Thirty adult patients, 18-70 years of either gender, undergoing cardiac surgery with CPB were enrolled. All patients received a reusable nasopharyngeal probe for monitoring core temperatures. In addition, the oesophageal temperatures were monitored with the TOE probe. The arterial outlet temperatures at the membrane oxygenator were also monitored and taken as the reference standard. Monitoring was performed every 5 min until 20 min, and then at 30 min during both the cooling and rewarming periods.
RESULTS
During cooling, the oesophageal and nasopharyngeal temperatures lagged behind the arterial outlet temperatures. However, the intra-class correlation of the oesophageal temperatures with the arterial outlet temperatures was better (range 0.58-0.74) than the correlation of the nasopharyngeal temperatures with the arterial outlet temperatures (range 0.46-0.62). During rewarming, the performance of the TOE probe was significantly superior to the nasopharyngeal probe. After 15 and 20 min of rewarming, there was a difference of ∼1°C between the oesophageal and nasopharyngeal temperatures. At 30 min of rewarming, the oesophageal and the arterial outlet temperatures were similar, while the nasopharyngeal temperatures still lagged by 0.5°C. Bias was significantly less both during cooling and warming between the oesophageal temperatures and arterial outlet temperatures.
CONCLUSIONS
Performance of the TOE probe as an oesophageal temperature probe is superior to the nasopharyngeal probe during CPB.
CLINICAL TRIAL REGISTRATION NUMBER: CTRI no 2020/10/028228; ctri.nic.in.