Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice.

Halkiadakis Ioannis, Patsea Eleni, Chatzimichali Katerina, Skouriotis Sotiris, Chalkidou Sonia, Amariotakis Georgios, Papakonstadinou Dimitrios, Theodossiadis George, Amariotakis Apostolos, Georgopoulos Gerasimos

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To compare intraocular pressure (IOP) readings taken using dynamic contour tonometry (DCT) with IOP readings taken with Goldmann applanation tonometry (GAT) in eyes with glaucoma or ocular hypertension.

METHODS

The present study included 100 eyes in 100 patients with glaucoma or ocular hypertension. After pachymetry DCT and GAT were performed. Intraocular pressures as measured with DCT and GAT were compared with one another and with central corneal thickness (CCT).

RESULTS

Mean DCT IOP measurements (20.1 +/- 4.3 mmHg) were significantly (p < 0.001) higher than GAT IOP values (17.9 +/- 4.7 mmHg). The mean difference between DCT and GAT measurements was 2.1 mmHg (range -3.4 to 9.7 mmHg). The difference followed a normal distribution. Measurements made with DCT and GAT correlated significantly with one another (Spearman's rho = 0.761, p < 0.001). Neither GAT nor DCT measurements showed a significant correlation with CCT (537 +/- 39 microm, range 458-656 microm). Multivariate regression analysis has shown that the difference between DCT and GAT is influenced significantly by ocular pulse amplitude (r = -0.334, p = 0.001) and it is not influenced by CCT (r = -0.106, p = 0.292).

CONCLUSIONS

In eyes with glaucoma or ocular hypertension, DCT facilitates suitable and reliable IOP measurements which are in good concordance with GAT readings. Variation in CCT cannot by itself explain the differences in measurements taken with DCT and GAT in a number of eyes.

Для цитирования: Halkiadakis Ioannis, Patsea Eleni, Chatzimichali Katerina, Skouriotis Sotiris, Chalkidou Sonia, Amariotakis Georgios, Papakonstadinou Dimitrios, Theodossiadis George, Amariotakis Apostolos, Georgopoulos Gerasimos. Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice.. Acta ophthalmologica. 2008. DOI: 10.1111/j.1755-3768.2008.01239.x.