A comparison study between the results of the Siemens ADVIA 120 analyzer and the manual method for differential leukocyte counts in sheep.

Oikonomidis Ioannis L, Brozos Christos, Kiossis Evangelos, Tsouloufi Theodora K, Kritsepi-Konstantinou Maria

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The ADVIA 120 is a widely used hematology analyzer, which has not been previously validated for determining differential leukocyte counts in sheep.

OBJECTIVES

We aimed to compare differential leukocyte counts on the ADVIA 120 (A-Diff) with counts obtained using the manual method (M-Diff) in sheep.

METHODS

Ethylenediaminetetraacetic acid-anticoagulated blood samples analyzed within 4 hours of collection were used. Samples with inappropriately filled tubes, detectable clots, overtly erroneous ADVIA peroxidase cytograms, and poor-quality blood smears were excluded from the study. Two independent observers compared the results of the A-Diff with those of the M-Diff. The M-diff was performed by counting 200 leukocytes on a blood smear.

RESULTS

Overall, 88 samples (44 rams and 44 ewes) were included. The correlation between the A-Diff and M-Diff was high for neutrophils (r = .873, P < .001), lymphocytes (r = .863, P < .001), and eosinophils (r = .750, P < .001), and low for monocytes (r = .212, P = .048). The Passing-Bablok regression analyses revealed constant error for eosinophils [1.17%; 95% confidence interval (CI): 0.67%, 1.55%] and proportional error for lymphocytes (0.84; 95% CI: 0.74, 0.95) and eosinophils (0.85; 95% CI: 0.74, 0.96). The Bland-Altman analyses revealed negative biases of 2.4% and 3.0% for neutrophils and lymphocytes, respectively, and positive biases of 3.2% and 0.8% for monocytes and eosinophils, respectively. The extremely low basophil percentages precluded a meaningful method comparison for this leukocyte type.

CONCLUSIONS

The ADVIA 120 appears to perform well for determining neutrophil, lymphocyte, and eosinophil percentages in sheep as compared with the manual method. However, blood smear examinations are recommended for the confirmation of abnormal differential leukocyte count results.

Для цитирования: Oikonomidis Ioannis L, Brozos Christos, Kiossis Evangelos, Tsouloufi Theodora K, Kritsepi-Konstantinou Maria. A comparison study between the results of the Siemens ADVIA 120 analyzer and the manual method for differential leukocyte counts in sheep.. Veterinary clinical pathology. 2021. DOI: 10.1111/vcp.12935.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.