Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Quantification of Four Endocannabinoids and Endocannabinoid-Like Substances in Plasma: Application in an HIV-Hepatitis C Virus Coinfected Population.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Gish Alexandr, Radwan Eqbal, Richeval Camille, Protopopescu Camelia, Wiart Jean-François, Hakim Florian, Allorge Delphine, Carrieri Patrizia, Barré Tangui, Gaulier Jean-Michel
Аннотация
BACKGROUND: The endocannabinoid system regulates immune function, inflammation, and metabolism and has gained increasing attention in clinical research. However, the instability, low abundance, and physicochemical complexity of endocannabinoids (ECs) and EC-like substances (EC-like) make their quantification in plasma analytically challenging.
OBJECTIVE AND METHODS: The authors developed and validated a liquid chromatography-tandem mass spectrometry method for the simultaneous quantification of N-arachidonoylethanolamide (AEA), 2-arachidonoylglycerol, and its isomer 1-arachidonoylglycerol (2-AG/1-AG), oleoylethanolamide (OEA), and palmitoylethanolamide (PEA) in human plasma.
RESULTS: Extraction was performed using liquid-liquid extraction combined with liquid chromatography-tandem mass spectrometry. The method achieved lower limits of quantification of 0.1 mcg/L for AEA, 0.2 mcg/L for 2-AG/1-AG, and 0.5 mcg/L for OEA and PEA, with calibration curves showing high linearity (R 2 ≥ 0.995). Intra-assay and interassay accuracy and precision were both within 15%. Additional validation parameters, including selectivity, recovery, carryover, matrix suitability, and dilution integrity, fulfilled regulatory criteria. Preanalytical handling significantly influenced concentrations: Delayed centrifugation and postthaw storage increased AEA, OEA, and PEA levels. The validated method was applied to plasma samples from patients coinfected with HIV and hepatitis C virus, enabling reproducible quantification of ECs and EC-like substances in a clinically relevant cohort.
CONCLUSIONS: The study demonstrates that biological and technical factors markedly affect plasma EC and EC-like concentrations. Standardized preanalytical processing is therefore essential for accurate measurement, and the proposed method provides a robust tool for clinical and pharmacological research.
OBJECTIVE AND METHODS: The authors developed and validated a liquid chromatography-tandem mass spectrometry method for the simultaneous quantification of N-arachidonoylethanolamide (AEA), 2-arachidonoylglycerol, and its isomer 1-arachidonoylglycerol (2-AG/1-AG), oleoylethanolamide (OEA), and palmitoylethanolamide (PEA) in human plasma.
RESULTS: Extraction was performed using liquid-liquid extraction combined with liquid chromatography-tandem mass spectrometry. The method achieved lower limits of quantification of 0.1 mcg/L for AEA, 0.2 mcg/L for 2-AG/1-AG, and 0.5 mcg/L for OEA and PEA, with calibration curves showing high linearity (R 2 ≥ 0.995). Intra-assay and interassay accuracy and precision were both within 15%. Additional validation parameters, including selectivity, recovery, carryover, matrix suitability, and dilution integrity, fulfilled regulatory criteria. Preanalytical handling significantly influenced concentrations: Delayed centrifugation and postthaw storage increased AEA, OEA, and PEA levels. The validated method was applied to plasma samples from patients coinfected with HIV and hepatitis C virus, enabling reproducible quantification of ECs and EC-like substances in a clinically relevant cohort.
CONCLUSIONS: The study demonstrates that biological and technical factors markedly affect plasma EC and EC-like concentrations. Standardized preanalytical processing is therefore essential for accurate measurement, and the proposed method provides a robust tool for clinical and pharmacological research.