MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1651-2227.2007.00631.x

Медицинская библиотека · Acta paediatrica (Oslo, Norway : 1992) · 2008

Standardized bench method for evaluating the efficacy of phototherapy devices.

Vreman H J, Wong R J, Murdock J R, Stevenson D K

Оценочное исследованиеСтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

AIM: As phototherapy (PT) devices employ a variety of broadband light sources, we developed and tested a standardized bench method for evaluating the efficacy of some devices.

METHODS

To evaluate efficacy, we quantified the in vitro photodegradation rate (expressed as t1/2) of unconjugated bilirubin in solution at 37 degrees C during exposure to a given light source at its mean delivered irradiance to the 2D body surface area (BSA) of newborn models. Reproducibility (between-day variation) of the method was determined at irradiance levels from 10 to 70 microW/cm2/nm on three different days.

RESULTS

Between-day t1/2 measurements had coefficients of variation from 3% to 10%. When t1/2 values were normalized to the exposable 2D horizontal BSA, halogen lamp devices, without and with fiberoptics, were least effective (t1/2=60-108 min and 100-126 min for preterm and term models, respectively). Fluorescent tube devices had t1/2=19-78 min and 25-78 min, for preterm and term models, respectively. Light-emitting diode (LED)-based devices yielded the shortest t1/2 values (16-24 min) for preterm and term newborn models.

CONCLUSIONS

We demonstrated the applicability of the method through the determination of the efficacy of several commercially available PT devices. This standardized method is reproducible and effectively evaluates the relative in vitro efficacy of various devices and may guide further in vitro and in vivo evaluations of devices.

Для цитирования: Vreman H J, Wong R J, Murdock J R, Stevenson D K. Standardized bench method for evaluating the efficacy of phototherapy devices.. Acta paediatrica (Oslo, Norway : 1992). 2008. DOI: 10.1111/j.1651-2227.2007.00631.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.