In-vitro comparison of mechanical resistance between two straight plates and a Y-plate for fixation of mandibular condyle fractures.
Аннотация
OBJECTIVE
To compare the mechanical resistance of conventional plates with a modified Y-plate in mandibular condyle fracture.
MATERIALS AND
METHODS
40 synthetic hemi-mandibular polyurethane replicates were used. Two groups of fixing materials were included: 20 mandibles fixed with 40 straight plates with four roles each (2.0 mm system) and 20 mandibles fixed with 20 Y-shaped plates (2.0 mm system and 1.5 mm thickness). The samples were submitted to linear loading in the following directions: 10 mandibles from each group from lateral to medial and 10 mandibles from each group from anterior to posterior. A universal loading machine, Instron Universal 4411, was used to performed the tests, and the resistance of each fixed mandible was assessed at displacements of 1 mm, 3 mm, and 5 mm. The t-test for independent samples was performed. The p value was set at <0.05.
RESULTS
The worst mechanical resistance was found in the Y-plates loaded from anterior to posterior. The best mechanical resistance was in two straight hole plates loaded from lateral to medial. The Y-plates and two straight hole plates showed similar resistance when a lateral to medial load was applied.
CONCLUSION
There was no difference between the plates when lateral to medial loading was tested.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.