Osteolytic changes accompanying degradation of absorbable fracture fixation implants.

Böstman O M

СтатьяПеревод в очереди

Аннотация

We analysed the radiographs of 67 patients with displaced malleolar fractures treated by open reduction and internal fixation using absorbable polyglycolide rods. Seventeen patients developed a discharging inflammatory foreign-body reaction, a complication unique to these fixation devices. In 34 patients ovoid osteolytic foci, usually 5 to 10 mm in diameter, appeared within the implant channels six to 12 weeks after the operation. The same lesion occurred in 14 of the 17 patients who developed a foreign-body reaction, whereas only 20 out of the 50 patients with an uneventful course showed osteolytic areas (p less than 0.01). In patients with a foreign-body reaction the osteolytic foci tended to occur in the deepest parts of the implant channels. However, after one year the normal structure of the bone was restored.

Для цитирования: Böstman O M. Osteolytic changes accompanying degradation of absorbable fracture fixation implants.. The Journal of bone and joint surgery. British volume. 1991. DOI: 10.1302/0301-620x.73b4.1649195.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.