MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1479-828x.2010.01166.x

Медицинская библиотека · The Australian & New Zealand journal of obstetrics & gynaecology · 2010

A randomised controlled trial of two instruments for vacuum-assisted delivery (Vacca Re-Usable OmniCup and the Bird anterior and posterior cups) to compare failure rates, safety and use effectiveness.

Mola Glen D L, Kuk Joseph M

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Most previous trials of vacuum-assisted delivery have been in settings with high rates of instrumental vaginal delivery (8-12%) and high rates of failure to deliver with the intended instrument (20-30%). Over the past 20 years, vacuum-assisted delivery rates at the Port Moresby General Hospital have been 3-4% with failure rates of <3%.

OBJECTIVE

The objective is to compare the failure rates of two vacuum extractor instruments, the Vacca Re-Usable Omnicup and the Bird Vacuum delivery system (anterior and posterior cups).

SETTING

Port Moresby General national referral and teaching Hospital (PMGH), Papua New Guinea.

POPULATION: Two hundred consecutive women requiring assisted delivery, June-December, 2007.

METHODS

When a woman required an assisted delivery, she was randomised into either the Vacca Re-Usable Omnicup (Clinical Innovations Inc.) or Bird anterior or posterior metal cup (depending upon the position of the vertex). One hundred women were randomised to each vacuum device. Statistical analysis was on 'an intention-to-treat' basis.

MAIN OUTCOME MEASURES: The main outcome measure was the successful completion of the delivery with the allocated instrument. Secondary outcomes were maternal trauma (episiotomy and trauma to the maternal genital tract), significant scalp trauma (sub-galeal haemorrhage or serious abrasion) and fetal and neonatal outcomes (Apgar score less than seven at 5 minutes, days spent in the Special Care Nursery and neonatal death).

RESULTS

Failure rates for both Omnicup (2/100) and Bird metal cups (6/100) were not statistically different (RR 1.05, 95% CI 0.99-1.12; P = 0.17). Rates of maternal trauma and fetal scalp trauma were similar in both groups.

CONCLUSION

Both the Vacca re-useable Omnicup and the Bird metal cups are very effective instruments to achieve successful assisted delivery and equally so. Failures and problems were associated with not applying the vacuum cup to the flexion point on the fetal scalp and the mechanical faults with vacuum equipment devices.

Для цитирования: Mola Glen D L, Kuk Joseph M. A randomised controlled trial of two instruments for vacuum-assisted delivery (Vacca Re-Usable OmniCup and the Bird anterior and posterior cups) to compare failure rates, safety and use effectiveness.. The Australian & New Zealand journal of obstetrics & gynaecology. 2010. DOI: 10.1111/j.1479-828x.2010.01166.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.